A file that fails goes back to the lab
The validation gate
Pick a file, or bring your own
File: Chemistry results from the old Navify export, with its errors (Results messages from the chemistry analysers, read on the blood chemistry lane)
Quarantined
At least one check failed, so nothing from this file goes on: not even the rows that passed.
12 rows read · 5 passed every check · 0 rows go on to review · 12 problems stop the file · 1 question for the reviewer
- 1
Check 1
Complete and readable
Every field VERA needs is there, no character is garbled and no identifier is cut short.
2 problems and 1 question in this file
- Message number missing or cut short. The message number “MSG-TRUNCA” is cut short, so this message cannot be traced.The lab sends the full message number with every message.
- Garbled characters. Creatinine: a character of the unit arrived garbled (“?mol/L”), so the unit cannot be read.The lab sends the file in standard text encoding, unchanged on the way, so the µ sign arrives intact.
- Instrument not named. The file does not name the instrument that measured the results.The lab names the instrument with the results: two instruments can give one test different ranges.
- 2
Check 2
Units
Every value has a unit VERA knows, fits its test, and is converted only through the laboratory's agreed conversion table.
4 problems in this file
- Range in another unit, result 1. Calcium: 2.44 mmol/L was judged against a range of 8.5–10.5 mg/dL, which is in another unit. In mg/dL the value is 9.8 mg/dL, so its high or low flag cannot be trusted.The lab converts the value and judges its flag against a range in the same unit: a value, its unit and its range travel together.
- Range in another unit, result 4. Creatinine: 88 µmol/L was judged against a range of 0.70–1.30 mg/dL, which is in another unit. In mg/dL the value is 0.99 mg/dL, so its high or low flag cannot be trusted.The lab converts the value and judges its flag against a range in the same unit: a value, its unit and its range travel together.
- Range in another unit, result 5. Cortisone: 18 ng/mL was judged against a range of 0.5–2.5 µg/dL, which is in another unit. In µg/dL the value is 1.8 µg/dL, so its high or low flag cannot be trusted.The lab converts the value and judges its flag against a range in the same unit: a value, its unit and its range travel together.
- Unit missing, result 7. Glucose has no unit, so VERA cannot read its value 92.The lab sends the unit with every value.
- 3
Check 3
Known codes
Every test, microbe and gene code is one VERA can look up, so nothing is guessed.
1 problem in this file
- Unknown test code, result 6. The laboratory's code “FAKE-99” (Unknown Analyte) is not a test code VERA can look up, so VERA does not know what was measured.The lab supplies the standard test code. VERA never invents one, and adds a translation of the laboratory's own code only after clinical sign-off.
- 4
Check 4
Possible values
Every value is one a living body can have, and its high or low flag agrees with the value and its range.
4 problems in this file
- Flag disagrees with the range, result 1. Calcium is flagged low, but 9.8 mg/dL is inside its range of 8.5–10.5 mg/dL.The lab sets the flag again from the converted value and a range in the same unit: the flag and the value must agree.
- Impossible value, result 2. Total protein: 124.8 g/dL is not a value a living body can have (VERA accepts 3.0–10.0 g/dL).Most likely a wrong unit or a set-up error at the lab. The lab fixes it and sends the result again; VERA never changes a value.
- Flag disagrees with the range, result 4. Creatinine is flagged high, but 0.99 mg/dL is inside its range of 0.70–1.30 mg/dL.The lab sets the flag again from the converted value and a range in the same unit: the flag and the value must agree.
- Flag disagrees with the range, result 5. Cortisone is flagged very high, but 1.8 µg/dL is inside its range of 0.5–2.5 µg/dL.The lab sets the flag again from the converted value and a range in the same unit: the flag and the value must agree.
- 5
Check 5
Nothing missing or repeated
No result twice, every test the panel promises is there, and a genome file says which parts it covered.
1 problem in this file
- Same result twice, result 9. Calcium is sent twice: this is the same result as result 1.The lab sends each result once. VERA never picks between two copies.
The release rule
All or nothing
On this file: 5 rows passed on their own, and none goes on, because the file failed a check.
Open this file in the Ingestor

VERA™ refuses a file that fails and reports why; it never corrects one quietly
When a file fails, VERA never quietly corrects it. The report of what failed goes to the lab, the lab fixes its set-up at source, and sends the file again. Quiet fixes are how the calcium incident happened: a fix in one place broke another, and nobody could see where.
Trust re-verification
Status: Proposed QC ritualThe laboratory checks a share of results again by hand, on a fixed schedule
A random 5% of results checked again by hand, as a monthly quality check, with the share that agrees recorded as a measure of the lab-to-twin boundary: the laboratory's equivalent of the knowledge graph's build check.
And where they are not used
Where frontier models belong
Ingestion
No modelsFixed readers, signed tables, hard stops
Fusion
No modelsMatching codes to people inside a sealed, trusted computer
Interpretation
ModelsLinks across blood, genome and microbes, kept as research evidence
Presentation
ModelsPlain-language explanations of checked results only
Fail-closed release path
After this gate: the VERA™ release gates
These five checks judge a file. Before anything reaches the twin, the VERA release path runs its own five fail-closed gates. Any failed or unknown gate means quarantine: a visible reason code, no patient-graph write and no learning event.
- 01IdentityOrder, subject, specimen, aliquot and run agree.
- 02CompletenessEvery required native output and its reviewed context is present.
- 03Analytical QCInstrument and analysis QC pass; callers do not disagree.
- 04Semantic QCEvery value binds to a known code and unit; nothing is guessed.
- 05ReleaseAn authorised reviewer signs the release.