Hardware integration
VERA™ Edge on DGX Spark
Concept
Status: PlannedThe enclosure in the decks (external bracket, dual network, read-only on the instrument) is a concept render. Planned: a DGX Spark in the edge enclosure. An NVIDIA DGX Spark is installed on site beside the Oxford Nanopore PromethION 48 as the VERA Edge read-only sidecar. This page is the integration specification.
Interfaces
Ports
| Port | Interface | Rule |
|---|---|---|
| DATA IN | USB‑C / 10 GbE receive from the export share | Read‑only capture. No control packets toward the instrument. |
| UPLINK | RJ‑45 10 GbE | VERA Core zone only. Never shares a switch with the instrument LAN. |
| PWR | Official 240 W inlet | Isolated bench circuit, separate from the analyser PSU. |
| AUX I/O | USB‑C console, LEDs | PWR / LNK / DATA / SEC status. |
DATA IN
- Interface
- USB-C / 10 GbE receive from the export share
- Rule
- Read-only capture. No control packets toward the instrument.
UPLINK
- Interface
- RJ-45 10 GbE
- Rule
- VERA Core zone only. Never shares a switch with the instrument LAN.
PWR
- Interface
- Official 240 W inlet
- Rule
- Isolated bench circuit, separate from the analyser PSU.
AUX I/O
- Interface
- USB-C console, LEDs
- Rule
- PWR / LNK / DATA / SEC status.
Fail-closed: if identity, digest or uplink drops, Stage 1 does not write facts.
Nine steps, one afternoon
Commissioning
- 01Confirm the nameplate serial against PREPAIRE‑IRENA‑ASSET‑REGISTRY@3.0.0.
- 02Fit the external bracket. Do not open the analyser.
- 03Seat the Spark. Record the GB10 module identity.
- 04Point DATA IN at the instrument export share only.
- 05Connect the 10 GbE uplink to VERA Core, never the instrument LAN. Leave QSFP on that zone.
- 06Power from the isolated 240 W feed. Confirm PWR / LNK / DATA / SEC.
- 07Boot DGX OS, load the signed VERA Edge image, admit plugin digests.
- 08Run a physical control (NIST / Zymo / vendor QC) before any Cohort 50 sample.
- 09Seal one native run, confirm the SHA‑256, run Stage 1, inspect the bound facts.
Dual-network rule
The Spark reads from the instrument's network and never connects it to another network.
- The Spark is an end point on each network: IP forwarding off, separate VRFs or physically separate switches.
- ConnectX-7 QSFP stays on VERA Core for pairing Sparks, never on the analyser network.
- No path from the analysis sandbox back into the analyser control plane.
- Instrument write path
- DATA IN, read-only
- VERA Edge on the Spark
- UPLINK 10 GbE, optional ConnectX-7
- VERA Core zone
Mechanical and thermal
Mounted outside the analyser
- External bracket only. No fasteners into fluidics or optics; service can lift the node without opening the analyser.
- Exhaust away from the instrument intake.
- NVIDIA operating window: 5–30 °C, 10–90% RH non-condensing.
- The analyser does not carry the load: enclosure plus the 1.2 kg module sit on the bench.
Edge deck · product sheet
The concept, as the deck drew it

The sheet's “seven governed edge nodes” is the deck's concept, and the current ask is one DGX Spark evaluation node. Physical one-to-one Sparks are approved only after qualification (DEC-V3-004).