LPR-HMS-000001
Hamilton Microlab STAR
Use the validated VENUS REST/LIMS interface and archive method plus detailed run logs after every execution.
ActiveReady for the laboratory's validation

Extraction playbook
- Record instrument, deck layout, method checksum, VENUS version, operator, consumable lots, and carrier identities.
- Receive barcode and worklist identity from LIM through the authenticated integration.
- Capture executed commands, liquid handling events, volumes, errors, and timestamps.
- Hash and seal the method and complete run trace.
- Publish tube-to-well lineage only after volume, barcode, and error gates pass.
Connection to VERA™
Reads its log filesVERA reads a Hamilton VENUS trace file from this machine when it is handed over as a file, and maps each line to a step with its sample and plate.
Instrument identity
- Manufacturer
- Hamilton
- Model
- Microlab Star
- Serial
- ••5K
- Facility
- Library Preparation Room
- Evidence role
- pre-analytical
- Software mode
- EXTERNAL VENDOR RUNTIME
Evidence contract
- Native evidence
- method file, run trace, executed commands, liquid-level and volume events, errors, audit trail
- Open derivatives
- REST/LIMS events, run-event JSON, plate/tube lineage
- Vendor software
- VENUS
- KG eligibility
- Disabled pending dual site activation
Acceptance tests
Positive and negative identity fixtures produce the approved lineage graph.
AwaitedAspirate/dispense errors and manual interventions force review.
AwaitedMethod checksum mismatch blocks execution evidence release.
AwaitedPrimary sources
Site activation attestations
No site approval is recorded in your session. Patient run creation remains disabled.
VERA refuses an approval of this instrument today, as it would at the lab:
- It is not an analyser, so it cannot start a patient run.
A revocation can still be recorded. Try an approval to see the rule bite.