DemonstrationA sample campaign: invented records and invented staff. Every step can be tried on this page, and nothing is recorded. About the samples
Sysmex XR-1000: Full blood count, version 1
Open: evidence being gatheredThe campaign is open: evidence is still being recorded, and no closing decision exists. Opened 23 Nov 2026.
Next step: the evidence decision
DemonstrationVERA scores the records when the quality manager or laboratory director asks.
VERA scores the records and records the result. On today's records the campaign stays open afterwards.
A demonstration: the step is checked and answered by the rules of a laboratory's own VERA, on the sample's invented records. Nothing is recorded, and what you try is gone when you leave this page.
- Step 1DoneWhat was tested
The campaign names one analyser, one assay protocol, one release of VERA with its two parts of software, one pipeline and every clinical table.
Recorded when the campaign is opened, by the quality manager, laboratory director or platform administrator.
- Step 2Under wayThe evidence
8 records: 7 passed, 1 failed. More can be recorded while the campaign is open.
Recorded by a quality manager, laboratory scientist or laboratory director.
- Step 3NextThe evidence decision
Not recorded yet. The quality manager or the laboratory director asks VERA to score the evidence.
Scored by VERA when the quality manager or laboratory director asks.
- Step 4WaitingThe approval
Comes after a closing evidence decision.
Given by the laboratory director, and by nobody else.
Step 1 What was tested
This copy is outside it
Who: Recorded when the campaign is opened, by the quality manager, laboratory director or platform administrator.
The scope is exactly what was tested: one analyser, one assay protocol, one release of VERA and every clinical table it reads. Each item is named by a digest, a fingerprint of its content: change one value and the digest changes. An approval covers a copy of VERA only while every item is the same.
- Analyser
- Sysmex XR-1000Entry SRA-SXA-000001 of the asset register
- Assay protocol
- Full blood count, version 1
- Release of VERA
- vera-2026.11.29-4 (build bf941dc)This copy runs the same.
- Clinical tables
- 20 tablesEvery table on this copy has the content the scope names.
All 20 tables of the scope, each with its digest
| Table | What it decides | Digest of its content | Against this copy |
|---|---|---|---|
Analyser flags and interference limits | The releasable analyser flags and index limits | 631357e70edf… | Same as this copy |
Biological variation | The biological variation behind personal ranges | 794171435ee3… | Same as this copy |
Calculated indices | The calculated indices | cd8fed5cec3b… | Same as this copy |
Coded references | The coded references on the report | 2feb8209f959… | Same as this copy |
Critical values | The critical limits | fafc8b999ba1… | Same as this copy |
Critical-result escalation | The critical-call clock and escalation chain | 1fbdc424c734… | Same as this copy |
Delta checks | The delta checks | 4fb6a0dc7dc7… | Same as this copy |
Genotype starting ranges | The genotype-specific starting ranges of personal ranges | aed748ac6fe5… | Same as this copy |
KERNEL's code crosswalk | KERNEL's local codes to LOINC | d2ab641633be… | Same as this copy |
Laboratory details on the report | The laboratory facts the report prints | f2d8b5852f9e… | Same as this copy |
Limits of the intake checks | The parameters of the ingestion gate's five checks | c489448479fc… | Same as this copy |
Malaffi message profile | The Malaffi message profile | e68294f3b449… | Same as this copy |
Malaffi section map | The Malaffi section of each marker | 904ea78c91ec… | Same as this copy |
Order rules | The order rules and pre-analytical limits | 6fc79d6c9275… | Same as this copy |
QC rules and control targets | The QC rules, targets and calibration intervals | 1b81ca0d91fb… | Same as this copy |
Staff register | Who may authorise a report | 231e232e6166… | Same as this copy |
Standard ranges | The standard ranges by sex and age | fef8bcb6c604… | Same as this copy |
Sysmex host codes | The Sysmex host codes | fa77610b6bd5… | Same as this copy |
Test catalogue and code map | The catalogue: each marker's code, unit, specimen and tube (the code map and unit table) | 1a35793f1f79… | Same as this copy |
Unit conversions | The unit conversions | 8edb4694b586… | Same as this copy |
The software of the scope, each part with its digest
- Platform
sha256:201f7b5c491c…This copy runs the same.- Worker, which runs the pipelines
sha256:34287f72d5f8…Compared once this copy is told this part's digest, when it is installed for a laboratory.- Pipeline
90a115d2a75b…- Software that was tested
- Sample result packager, version sample
- List of test files
3ce340d21fed…
Step 2 The evidence
8 records: 7 passed, 1 failed
Who: Recorded by a quality manager, laboratory scientist or laboratory director.
Each record is one test file with a known expected result, run by one operator on one reagent lot at one site. It passes when the output agrees with the expected result. A record weighs more the further it is from a single run: another operator, another lot, another site, an outside sample.
| Kind of evidence | Weight of one record | Records | Passed |
|---|---|---|---|
One run | 1 | 6 | 5 |
Another operator | 1.25 | 2 | 2 |
Every record, with its result (8)
| Test file | Result | Kind | Operator, lot, site | Digest of the output | Recorded |
|---|---|---|---|---|---|
SAMPLE-FILE-001 | Passed | One run | sample-operator-1, sample-lot-A, sample-site-1 | bf483cce431d… | 23 Nov 2026, 09:47 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-002 | Passed | One run | sample-operator-1, sample-lot-A, sample-site-1 | b80b428cc174… | 23 Nov 2026, 10:34 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-003 | Passed | One run | sample-operator-1, sample-lot-A, sample-site-1 | d7509d381524… | 23 Nov 2026, 11:21 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-004 | Passed | One run | sample-operator-1, sample-lot-A, sample-site-1 | 9e5dc261d445… | 23 Nov 2026, 12:08 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-005 | Failed | One run | sample-operator-1, sample-lot-A, sample-site-1 | 5e395b08e354… | 23 Nov 2026, 12:55 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-006 | Passed | One run | sample-operator-1, sample-lot-A, sample-site-1 | 658295ff64ef… | 23 Nov 2026, 13:42 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-007 | Passed | Another operator | sample-operator-2, sample-lot-A, sample-site-1 | fe9b4df50e05… | 23 Nov 2026, 14:29 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-008 | Passed | Another operator | sample-operator-2, sample-lot-A, sample-site-1 | 983a21bcbae4… | 23 Nov 2026, 15:16 GST Sign-in Sample laboratory scientist |
Step 3 The evidence decision
No decision yet
Who: Scored by VERA when the quality manager or laboratory director asks.
Nobody types this decision in. VERA scores the records under the laboratory's evidence policy and records the score with its reason. A decision that the evidence is sufficient, or that it is not, closes the campaign.
If the decision were recorded now
Not enough evidence yet
There are not yet enough records to score. The campaign stays open for more evidence. Counted by the policy's weights, the records add up to 8.5; the policy scores a campaign once they add up to 10.
VERA's reason, were the decision recorded nowThe records do not yet add up to the number the laboratory's evidence policy asks for before a campaign is scored.
Mandatory checks of the software: 11 of 11 passed
- The analyser's original file is sealed before anything converts or reads it. Passed.
- The converted copy agrees with the original file, with nothing lost. Passed.
- The sample, the run, the analyser, the operator, the reagent lot and every file are identified. Passed.
- The versions of the software, the models, the references, the methods and the settings are fixed. Passed.
- Legal approval covers exactly how the software is packaged, where it runs, who uses it and how it runs. Passed.
- The build is signed, and its signature and origin are checked against the approved signer. Passed.
- The list of every component in the build is present and genuine, was scanned for known faults, and belongs to this exact build. Passed.
- The security review passed: the threats considered, known faults, secrets, the network and the controls while running. Passed.
- The approved test files pass: positive, negative, at the limits, and files that must fail. Passed.
- The intended clinical or research use is approved in writing. Passed.
- Going back to the earlier build, withdrawing this build and repeating a result have each been shown. Passed.
Step 4 The approval
Open: evidence being gathered
Who: Given by the laboratory director, and by nobody else.
The evidence decision is a score. The approval is the laboratory director's own act: it names the evidence decision it approves, and only then may a clinical result measured in this scope be released.
No approval is recorded for this campaign.
The evidence decision comes first: this campaign has no closing decision yet.
What ends an approved scope
Today a clinical result of this analyser and assay protocol would not pass the scope check on this copy of VERA.
What VERA's scope check says today
- No campaign approved by the laboratory director covers this analyser and assay protocol (Full blood count, version 1).
- A changed clinical table
The scope names every clinical table by the digest of its content, so a copy on which one value of one table has changed is outside the scope until a campaign for the new content is approved.
- A new build
The scope names one numbered release and the two parts of the software built from it (the platform, and the worker that runs the pipelines), so a copy that runs any other release or any other build of a part is outside the scope until a campaign for that build is approved.
- A withdrawal
The laboratory director may withdraw the approval at any time, and it ends at once, also for a release already being prepared.
- A later rejection
When a later campaign of the same scope ends with the decision that the evidence is not sufficient, the earlier approval ends with it.
About this sample campaign
Demonstration
This copy of VERA is an open demonstration. This campaign is a sample: invented records and three invented members of staff, read, scored and compared by the code a laboratory's own VERA runs. The analysers, the clinical tables and their digests are this copy's own.
Two steps need signed-in people: the evidence decision (the quality manager or the laboratory director) and the approval of the scope (the laboratory director).
On this copy each step on this page can be tried as one of the invented members of staff (the sample laboratory scientist, the sample quality manager, the sample laboratory director). The answer comes from the same checks and the same scoring as on a laboratory's own VERA. What is tried is held in the page, sent with each step, and gone on leaving it.
On a laboratory's own VERA the approval is also checked against the signed staff register.
Who takes each step, and why nobody else may
- Recording the evidence
A quality manager, laboratory scientist or laboratory director, each signed in.
Each record is one test file run by one operator on one reagent lot at one site. The same record entered twice is refused, and it is scored once.
- The evidence decision
The quality manager or laboratory director, signed in.
Nobody types the decision in: VERA scores the records under the evidence policy and records the score with its reason. Any other sign-in is refused, and so is a request with no sign-in.
- Approving the scope
The laboratory director only, signed in and listed in the signed staff register, with a licence number, as valid on the day.
The quality manager cannot approve: the decision is a score, and the approval is the laboratory director's own act. A laboratory director who recorded evidence of the campaign cannot approve it either: a person does not approve tests they performed.
- Withdrawing an approval
The laboratory director, signed in.
A withdrawal names the approval it ends, with a reason. Both stay on record; nothing is changed or deleted.