Validation

VERA validation campaign

One campaign in four steps, from what was tested to the laboratory director's approval. The step that comes next is first on the page.

All validation campaigns

DemonstrationA sample campaign: invented records and invented staff. Every step can be tried on this page, and nothing is recorded. About the samples

Sysmex XR-1000: Full blood count, version 1

Open: evidence being gathered

The campaign is open: evidence is still being recorded, and no closing decision exists. Opened 23 Nov 2026.

Next step: the evidence decision

Demonstration

VERA scores the records when the quality manager or laboratory director asks.

VERA scores the records and records the result. On today's records the campaign stays open afterwards.

A demonstration: the step is checked and answered by the rules of a laboratory's own VERA, on the sample's invented records. Nothing is recorded, and what you try is gone when you leave this page.

  1. Step 1Done
    What was tested

    The campaign names one analyser, one assay protocol, one release of VERA with its two parts of software, one pipeline and every clinical table.

    Recorded when the campaign is opened, by the quality manager, laboratory director or platform administrator.

  2. Step 2Under way
    The evidence

    8 records: 7 passed, 1 failed. More can be recorded while the campaign is open.

    Recorded by a quality manager, laboratory scientist or laboratory director.

  3. Step 3Next
    The evidence decision

    Not recorded yet. The quality manager or the laboratory director asks VERA to score the evidence.

    Scored by VERA when the quality manager or laboratory director asks.

  4. Step 4Waiting
    The approval

    Comes after a closing evidence decision.

    Given by the laboratory director, and by nobody else.

Step 1 What was tested

This copy is outside it

Who: Recorded when the campaign is opened, by the quality manager, laboratory director or platform administrator.

The scope is exactly what was tested: one analyser, one assay protocol, one release of VERA and every clinical table it reads. Each item is named by a digest, a fingerprint of its content: change one value and the digest changes. An approval covers a copy of VERA only while every item is the same.

Analyser
Sysmex XR-1000Entry SRA-SXA-000001 of the asset register
Assay protocol
Full blood count, version 1
Release of VERA
vera-2026.11.29-4 (build bf941dc)This copy runs the same.
Clinical tables
20 tablesEvery table on this copy has the content the scope names.
All 20 tables of the scope, each with its digest
TableWhat it decidesDigest of its contentAgainst this copy
Analyser flags and interference limits
The releasable analyser flags and index limits631357e70edf…Same as this copy
Biological variation
The biological variation behind personal ranges794171435ee3…Same as this copy
Calculated indices
The calculated indicescd8fed5cec3b…Same as this copy
Coded references
The coded references on the report2feb8209f959…Same as this copy
Critical values
The critical limitsfafc8b999ba1…Same as this copy
Critical-result escalation
The critical-call clock and escalation chain1fbdc424c734…Same as this copy
Delta checks
The delta checks4fb6a0dc7dc7…Same as this copy
Genotype starting ranges
The genotype-specific starting ranges of personal rangesaed748ac6fe5…Same as this copy
KERNEL's code crosswalk
KERNEL's local codes to LOINCd2ab641633be…Same as this copy
Laboratory details on the report
The laboratory facts the report printsf2d8b5852f9e…Same as this copy
Limits of the intake checks
The parameters of the ingestion gate's five checksc489448479fc…Same as this copy
Malaffi message profile
The Malaffi message profilee68294f3b449…Same as this copy
Malaffi section map
The Malaffi section of each marker904ea78c91ec…Same as this copy
Order rules
The order rules and pre-analytical limits6fc79d6c9275…Same as this copy
QC rules and control targets
The QC rules, targets and calibration intervals1b81ca0d91fb…Same as this copy
Staff register
Who may authorise a report231e232e6166…Same as this copy
Standard ranges
The standard ranges by sex and agefef8bcb6c604…Same as this copy
Sysmex host codes
The Sysmex host codesfa77610b6bd5…Same as this copy
Test catalogue and code map
The catalogue: each marker's code, unit, specimen and tube (the code map and unit table)1a35793f1f79…Same as this copy
Unit conversions
The unit conversions8edb4694b586…Same as this copy
The software of the scope, each part with its digest
Platform
sha256:201f7b5c491c…This copy runs the same.
Worker, which runs the pipelines
sha256:34287f72d5f8…Compared once this copy is told this part's digest, when it is installed for a laboratory.
Pipeline
90a115d2a75b…
Software that was tested
Sample result packager, version sample
List of test files
3ce340d21fed…

Step 2 The evidence

8 records: 7 passed, 1 failed

Who: Recorded by a quality manager, laboratory scientist or laboratory director.

Each record is one test file with a known expected result, run by one operator on one reagent lot at one site. It passes when the output agrees with the expected result. A record weighs more the further it is from a single run: another operator, another lot, another site, an outside sample.

Kind of evidenceWeight of one recordRecordsPassed
One run
165
Another operator
1.2522
Every record, with its result (8)
Test fileResultKindOperator, lot, siteDigest of the outputRecorded
SAMPLE-FILE-001
PassedOne runsample-operator-1, sample-lot-A, sample-site-1bf483cce431d…23 Nov 2026, 09:47 GST
Sign-in Sample laboratory scientist
SAMPLE-FILE-002
PassedOne runsample-operator-1, sample-lot-A, sample-site-1b80b428cc174…23 Nov 2026, 10:34 GST
Sign-in Sample laboratory scientist
SAMPLE-FILE-003
PassedOne runsample-operator-1, sample-lot-A, sample-site-1d7509d381524…23 Nov 2026, 11:21 GST
Sign-in Sample laboratory scientist
SAMPLE-FILE-004
PassedOne runsample-operator-1, sample-lot-A, sample-site-19e5dc261d445…23 Nov 2026, 12:08 GST
Sign-in Sample laboratory scientist
SAMPLE-FILE-005
FailedOne runsample-operator-1, sample-lot-A, sample-site-15e395b08e354…23 Nov 2026, 12:55 GST
Sign-in Sample laboratory scientist
SAMPLE-FILE-006
PassedOne runsample-operator-1, sample-lot-A, sample-site-1658295ff64ef…23 Nov 2026, 13:42 GST
Sign-in Sample laboratory scientist
SAMPLE-FILE-007
PassedAnother operatorsample-operator-2, sample-lot-A, sample-site-1fe9b4df50e05…23 Nov 2026, 14:29 GST
Sign-in Sample laboratory scientist
SAMPLE-FILE-008
PassedAnother operatorsample-operator-2, sample-lot-A, sample-site-1983a21bcbae4…23 Nov 2026, 15:16 GST
Sign-in Sample laboratory scientist

Step 3 The evidence decision

No decision yet

Who: Scored by VERA when the quality manager or laboratory director asks.

Nobody types this decision in. VERA scores the records under the laboratory's evidence policy and records the score with its reason. A decision that the evidence is sufficient, or that it is not, closes the campaign.

If the decision were recorded now

Not enough evidence yet

There are not yet enough records to score. The campaign stays open for more evidence. Counted by the policy's weights, the records add up to 8.5; the policy scores a campaign once they add up to 10.

VERA's reason, were the decision recorded nowThe records do not yet add up to the number the laboratory's evidence policy asks for before a campaign is scored.

Mandatory checks of the software: 11 of 11 passed
  • The analyser's original file is sealed before anything converts or reads it. Passed.
  • The converted copy agrees with the original file, with nothing lost. Passed.
  • The sample, the run, the analyser, the operator, the reagent lot and every file are identified. Passed.
  • The versions of the software, the models, the references, the methods and the settings are fixed. Passed.
  • Legal approval covers exactly how the software is packaged, where it runs, who uses it and how it runs. Passed.
  • The build is signed, and its signature and origin are checked against the approved signer. Passed.
  • The list of every component in the build is present and genuine, was scanned for known faults, and belongs to this exact build. Passed.
  • The security review passed: the threats considered, known faults, secrets, the network and the controls while running. Passed.
  • The approved test files pass: positive, negative, at the limits, and files that must fail. Passed.
  • The intended clinical or research use is approved in writing. Passed.
  • Going back to the earlier build, withdrawing this build and repeating a result have each been shown. Passed.

Step 4 The approval

Open: evidence being gathered

Who: Given by the laboratory director, and by nobody else.

The evidence decision is a score. The approval is the laboratory director's own act: it names the evidence decision it approves, and only then may a clinical result measured in this scope be released.

No approval is recorded for this campaign.

The evidence decision comes first: this campaign has no closing decision yet.

What ends an approved scope

Today a clinical result of this analyser and assay protocol would not pass the scope check on this copy of VERA.

What VERA's scope check says today
  • No campaign approved by the laboratory director covers this analyser and assay protocol (Full blood count, version 1).
  • A changed clinical table

    The scope names every clinical table by the digest of its content, so a copy on which one value of one table has changed is outside the scope until a campaign for the new content is approved.

  • A new build

    The scope names one numbered release and the two parts of the software built from it (the platform, and the worker that runs the pipelines), so a copy that runs any other release or any other build of a part is outside the scope until a campaign for that build is approved.

  • A withdrawal

    The laboratory director may withdraw the approval at any time, and it ends at once, also for a release already being prepared.

  • A later rejection

    When a later campaign of the same scope ends with the decision that the evidence is not sufficient, the earlier approval ends with it.

About this sample campaign

Demonstration

This copy of VERA is an open demonstration. This campaign is a sample: invented records and three invented members of staff, read, scored and compared by the code a laboratory's own VERA runs. The analysers, the clinical tables and their digests are this copy's own.

Two steps need signed-in people: the evidence decision (the quality manager or the laboratory director) and the approval of the scope (the laboratory director).

On this copy each step on this page can be tried as one of the invented members of staff (the sample laboratory scientist, the sample quality manager, the sample laboratory director). The answer comes from the same checks and the same scoring as on a laboratory's own VERA. What is tried is held in the page, sent with each step, and gone on leaving it.

On a laboratory's own VERA the approval is also checked against the signed staff register.

Who takes each step, and why nobody else may

  • Recording the evidence

    A quality manager, laboratory scientist or laboratory director, each signed in.

    Each record is one test file run by one operator on one reagent lot at one site. The same record entered twice is refused, and it is scored once.

  • The evidence decision

    The quality manager or laboratory director, signed in.

    Nobody types the decision in: VERA scores the records under the evidence policy and records the score with its reason. Any other sign-in is refused, and so is a request with no sign-in.

  • Approving the scope

    The laboratory director only, signed in and listed in the signed staff register, with a licence number, as valid on the day.

    The quality manager cannot approve: the decision is a score, and the approval is the laboratory director's own act. A laboratory director who recorded evidence of the campaign cannot approve it either: a person does not approve tests they performed.

  • Withdrawing an approval

    The laboratory director, signed in.

    A withdrawal names the approval it ends, with a reason. Both stay on record; nothing is changed or deleted.

Demonstration copy