VERA™ validation campaign
DemonstrationA sample campaign: invented records and invented staff. Every step can be tried on this page, and nothing is recorded. About the samples
Roche cobas pure (c 303 and e 402): Clinical chemistry, version 1
Waiting for the laboratory directorVERA's evidence decision closed the campaign. The scope releases nothing until the laboratory director approves it. Opened 16 Nov 2026, closed 20 Nov 2026.
Next step: the approval of the scope
DemonstrationThe laboratory director takes this step, and nobody else can.
You approve the latest evidence decision for exactly the scope this campaign recorded. VERA checks the staff register before it records the approval.
The button opens once both entries are filled in.
On a laboratory's own VERA the approval also needs the signed staff register to list the laboratory director, with a licence number, as valid on the day.
A demonstration: the step is checked and answered by the rules of a laboratory's own VERA, on the sample's invented records. Nothing is recorded, and what you try is gone when you leave this page.
- Step 1DoneWhat was tested
The campaign names one analyser, one assay protocol, one release of VERA with its two parts of software, one pipeline and every clinical table.
Recorded when the campaign is opened, by the quality manager, laboratory director or platform administrator.
- Step 2DoneThe evidence
42 records: 42 passed, 0 failed.
Recorded by a quality manager, laboratory scientist or laboratory director.
- Step 3DoneThe evidence decision
Sufficient for monitored use, recorded on 20 Nov 2026.
Scored by VERA when the quality manager or laboratory director asks.
- Step 4NextThe approval
Waiting for the laboratory director.
Given by the laboratory director, and by nobody else.
Step 1 What was tested
This copy is outside it
Who: Recorded when the campaign is opened, by the quality manager, laboratory director or platform administrator.
The scope is exactly what was tested: one analyser, one assay protocol, one release of VERA and every clinical table it reads. Each item is named by a digest, a fingerprint of its content: change one value and the digest changes. An approval covers a copy of VERA only while every item is the same.
- Analyser
- Roche cobas pure (c 303 and e 402)Entry CP-CP-000001 of the asset register
- Assay protocol
- Clinical chemistry, version 1
- Release of VERA
- vera-2026.11.29-4 (build bf941dc)This copy runs the same.
- Clinical tables
- 20 tablesEvery table on this copy has the content the scope names.
All 20 tables of the scope, each with its digest
| Table | What it decides | Digest of its content | Against this copy |
|---|---|---|---|
Analyser flags and interference limits | The releasable analyser flags and index limits | 631357e70edf… | Same as this copy |
Biological variation | The biological variation behind personal ranges | 794171435ee3… | Same as this copy |
Calculated indices | The calculated indices | cd8fed5cec3b… | Same as this copy |
Coded references | The coded references on the report | 2feb8209f959… | Same as this copy |
Critical values | The critical limits | fafc8b999ba1… | Same as this copy |
Critical-result escalation | The critical-call clock and escalation chain | 1fbdc424c734… | Same as this copy |
Delta checks | The delta checks | 4fb6a0dc7dc7… | Same as this copy |
Genotype starting ranges | The genotype-specific starting ranges of personal ranges | aed748ac6fe5… | Same as this copy |
KERNEL's code crosswalk | KERNEL's local codes to LOINC | d2ab641633be… | Same as this copy |
Laboratory details on the report | The laboratory facts the report prints | f2d8b5852f9e… | Same as this copy |
Limits of the intake checks | The parameters of the ingestion gate's five checks | c489448479fc… | Same as this copy |
Malaffi message profile | The Malaffi message profile | e68294f3b449… | Same as this copy |
Malaffi section map | The Malaffi section of each marker | 904ea78c91ec… | Same as this copy |
Order rules | The order rules and pre-analytical limits | 6fc79d6c9275… | Same as this copy |
QC rules and control targets | The QC rules, targets and calibration intervals | 1b81ca0d91fb… | Same as this copy |
Staff register | Who may authorise a report | 231e232e6166… | Same as this copy |
Standard ranges | The standard ranges by sex and age | fef8bcb6c604… | Same as this copy |
Sysmex host codes | The Sysmex host codes | fa77610b6bd5… | Same as this copy |
Test catalogue and code map | The catalogue: each marker's code, unit, specimen and tube (the code map and unit table) | 1a35793f1f79… | Same as this copy |
Unit conversions | The unit conversions | 8edb4694b586… | Same as this copy |
The software of the scope, each part with its digest
- Platform
sha256:201f7b5c491c…This copy runs the same.- Worker, which runs the pipelines
sha256:34287f72d5f8…Compared once this copy is told this part's digest, when it is installed for a laboratory.- Pipeline
ad2eea98f53b…- Software that was tested
- Sample result packager, version sample
- List of test files
d24abd40238b…
Step 2 The evidence
42 records: 42 passed, 0 failed
Who: Recorded by a quality manager, laboratory scientist or laboratory director.
Each record is one test file with a known expected result, run by one operator on one reagent lot at one site. It passes when the output agrees with the expected result. A record weighs more the further it is from a single run: another operator, another lot, another site, an outside sample.
| Kind of evidence | Weight of one record | Records | Passed |
|---|---|---|---|
One run | 1 | 10 | 10 |
Another operator | 1.25 | 8 | 8 |
Another reagent lot | 1.5 | 8 | 8 |
Another site | 2 | 8 | 8 |
External proficiency sample | 3 | 8 | 8 |
Every record, with its result (42)
| Test file | Result | Kind | Operator, lot, site | Digest of the output | Recorded |
|---|---|---|---|---|---|
SAMPLE-FILE-001 | Passed | One run | sample-operator-1, sample-lot-A, sample-site-1 | 28be557e6334… | 16 Nov 2026, 09:47 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-002 | Passed | One run | sample-operator-1, sample-lot-A, sample-site-1 | 6057caaa154d… | 16 Nov 2026, 10:34 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-003 | Passed | One run | sample-operator-1, sample-lot-A, sample-site-1 | 9d929d04efb1… | 16 Nov 2026, 11:21 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-004 | Passed | One run | sample-operator-1, sample-lot-A, sample-site-1 | 92a7c5a9e3d1… | 16 Nov 2026, 12:08 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-005 | Passed | One run | sample-operator-1, sample-lot-A, sample-site-1 | 745efd073099… | 16 Nov 2026, 12:55 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-006 | Passed | One run | sample-operator-1, sample-lot-A, sample-site-1 | 5f2e40cd1146… | 16 Nov 2026, 13:42 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-007 | Passed | One run | sample-operator-1, sample-lot-A, sample-site-1 | 07b1165afbc0… | 16 Nov 2026, 14:29 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-008 | Passed | One run | sample-operator-1, sample-lot-A, sample-site-1 | 2849e20b1b65… | 16 Nov 2026, 15:16 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-009 | Passed | One run | sample-operator-1, sample-lot-A, sample-site-1 | 8dad7e1217b0… | 16 Nov 2026, 16:03 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-010 | Passed | One run | sample-operator-1, sample-lot-A, sample-site-1 | 709bd297f572… | 16 Nov 2026, 16:50 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-011 | Passed | Another operator | sample-operator-2, sample-lot-A, sample-site-1 | 2ee316ab9cba… | 16 Nov 2026, 17:37 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-012 | Passed | Another operator | sample-operator-2, sample-lot-A, sample-site-1 | 52da6bfdf61d… | 16 Nov 2026, 18:24 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-013 | Passed | Another operator | sample-operator-2, sample-lot-A, sample-site-1 | b103fbc55bdc… | 16 Nov 2026, 19:11 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-014 | Passed | Another operator | sample-operator-2, sample-lot-A, sample-site-1 | cb381f7245bf… | 16 Nov 2026, 19:58 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-015 | Passed | Another operator | sample-operator-2, sample-lot-A, sample-site-1 | 302e78bf3046… | 16 Nov 2026, 20:45 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-016 | Passed | Another operator | sample-operator-2, sample-lot-A, sample-site-1 | b6a881c3eb85… | 16 Nov 2026, 21:32 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-017 | Passed | Another operator | sample-operator-2, sample-lot-A, sample-site-1 | 0235a9a4ff10… | 16 Nov 2026, 22:19 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-018 | Passed | Another operator | sample-operator-2, sample-lot-A, sample-site-1 | eb1c6e9d4178… | 16 Nov 2026, 23:06 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-019 | Passed | Another reagent lot | sample-operator-1, sample-lot-B, sample-site-1 | 745dc0d1e339… | 16 Nov 2026, 23:53 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-020 | Passed | Another reagent lot | sample-operator-1, sample-lot-B, sample-site-1 | f4555221ccb9… | 17 Nov 2026, 00:40 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-021 | Passed | Another reagent lot | sample-operator-1, sample-lot-B, sample-site-1 | 49672196b6eb… | 17 Nov 2026, 01:27 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-022 | Passed | Another reagent lot | sample-operator-1, sample-lot-B, sample-site-1 | f3449f0e5009… | 17 Nov 2026, 02:14 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-023 | Passed | Another reagent lot | sample-operator-1, sample-lot-B, sample-site-1 | ee8cb75308bb… | 17 Nov 2026, 03:01 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-024 | Passed | Another reagent lot | sample-operator-1, sample-lot-B, sample-site-1 | a126fd26d6ee… | 17 Nov 2026, 03:48 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-025 | Passed | Another reagent lot | sample-operator-1, sample-lot-B, sample-site-1 | 6b4be42d158d… | 17 Nov 2026, 04:35 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-026 | Passed | Another reagent lot | sample-operator-1, sample-lot-B, sample-site-1 | 3ffcd28550d8… | 17 Nov 2026, 05:22 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-027 | Passed | Another site | sample-operator-1, sample-lot-A, sample-site-2 | 490df0b20259… | 17 Nov 2026, 06:09 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-028 | Passed | Another site | sample-operator-1, sample-lot-A, sample-site-2 | 966d695a3333… | 17 Nov 2026, 06:56 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-029 | Passed | Another site | sample-operator-1, sample-lot-A, sample-site-2 | e555b0d929dd… | 17 Nov 2026, 07:43 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-030 | Passed | Another site | sample-operator-1, sample-lot-A, sample-site-2 | 84985c308c7f… | 17 Nov 2026, 08:30 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-031 | Passed | Another site | sample-operator-1, sample-lot-A, sample-site-2 | d09687714a50… | 17 Nov 2026, 09:17 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-032 | Passed | Another site | sample-operator-1, sample-lot-A, sample-site-2 | 4bc6283aab72… | 17 Nov 2026, 10:04 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-033 | Passed | Another site | sample-operator-1, sample-lot-A, sample-site-2 | 6fe60f78743f… | 17 Nov 2026, 10:51 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-034 | Passed | Another site | sample-operator-1, sample-lot-A, sample-site-2 | b44a91348ac1… | 17 Nov 2026, 11:38 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-035 | Passed | External proficiency sample | sample-operator-1, sample-lot-A, sample-scheme | ffe601009e53… | 17 Nov 2026, 12:25 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-036 | Passed | External proficiency sample | sample-operator-1, sample-lot-A, sample-scheme | 1d863d2e6168… | 17 Nov 2026, 13:12 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-037 | Passed | External proficiency sample | sample-operator-1, sample-lot-A, sample-scheme | 2fe85cef3669… | 17 Nov 2026, 13:59 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-038 | Passed | External proficiency sample | sample-operator-1, sample-lot-A, sample-scheme | 10949bd1315b… | 17 Nov 2026, 14:46 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-039 | Passed | External proficiency sample | sample-operator-1, sample-lot-A, sample-scheme | a26406dc7dec… | 17 Nov 2026, 15:33 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-040 | Passed | External proficiency sample | sample-operator-1, sample-lot-A, sample-scheme | 975a624a996c… | 17 Nov 2026, 16:20 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-041 | Passed | External proficiency sample | sample-operator-1, sample-lot-A, sample-scheme | 27bc45c22685… | 17 Nov 2026, 17:07 GST Sign-in Sample laboratory scientist |
SAMPLE-FILE-042 | Passed | External proficiency sample | sample-operator-1, sample-lot-A, sample-scheme | 34fd6f77231a… | 17 Nov 2026, 17:54 GST Sign-in Sample laboratory scientist |
Step 3 The evidence decision
Sufficient for monitored use
Who: Scored by VERA when the quality manager or laboratory director asks.
Nobody types this decision in. VERA scores the records under the laboratory's evidence policy and records the score with its reason. A decision that the evidence is sufficient, or that it is not, closes the campaign.
Every mandatory check passed, and VERA's output agreed with the expected results often enough for the policy's limit for restricted, monitored use. Counted by the policy's weights, the records add up to 72. From them VERA works out how often its output agrees with the expected result: in at least 95.0% of runs. The figure is set low on purpose: the chance that the true share is lower is 2.5%. The policy asks for at least 95.0% for restricted, monitored use. For use without that restriction it asks for at least 99.0%, from records that add up to 30 or more.
Recorded 20 Nov 2026, 13:30 GST, under sign-in Sample quality manager, by version 1.0.0 of the evidence policy.
VERA's reasonEvery mandatory check passed, and the cautious estimate of how often VERA's output agrees with the expected result meets the policy's limit for restricted, monitored use.
Mandatory checks of the software: 11 of 11 passed
- The analyser's original file is sealed before anything converts or reads it. Passed.
- The converted copy agrees with the original file, with nothing lost. Passed.
- The sample, the run, the analyser, the operator, the reagent lot and every file are identified. Passed.
- The versions of the software, the models, the references, the methods and the settings are fixed. Passed.
- Legal approval covers exactly how the software is packaged, where it runs, who uses it and how it runs. Passed.
- The build is signed, and its signature and origin are checked against the approved signer. Passed.
- The list of every component in the build is present and genuine, was scanned for known faults, and belongs to this exact build. Passed.
- The security review passed: the threats considered, known faults, secrets, the network and the controls while running. Passed.
- The approved test files pass: positive, negative, at the limits, and files that must fail. Passed.
- The intended clinical or research use is approved in writing. Passed.
- Going back to the earlier build, withdrawing this build and repeating a result have each been shown. Passed.
The campaign is closed and takes no further record, so this page asks for no new score.
Step 4 The approval
Waiting for the laboratory director
Who: Given by the laboratory director, and by nobody else.
The evidence decision is a score. The approval is the laboratory director's own act: it names the evidence decision it approves, and only then may a clinical result measured in this scope be released.
No approval is recorded for this campaign.
What ends an approved scope
Today a clinical result of this analyser and assay protocol would not pass the scope check on this copy of VERA.
What VERA's scope check says today
- No campaign approved by the laboratory director covers this analyser and assay protocol (Clinical chemistry, version 1).
- A changed clinical table
The scope names every clinical table by the digest of its content, so a copy on which one value of one table has changed is outside the scope until a campaign for the new content is approved.
- A new build
The scope names one numbered release and the two parts of the software built from it (the platform, and the worker that runs the pipelines), so a copy that runs any other release or any other build of a part is outside the scope until a campaign for that build is approved.
- A withdrawal
The laboratory director may withdraw the approval at any time, and it ends at once, also for a release already being prepared.
- A later rejection
When a later campaign of the same scope ends with the decision that the evidence is not sufficient, the earlier approval ends with it.
About this sample campaign
Demonstration
This copy of VERA is an open demonstration. This campaign is a sample: invented records and three invented members of staff, read, scored and compared by the code a laboratory's own VERA runs. The analysers, the clinical tables and their digests are this copy's own.
Two steps need signed-in people: the evidence decision (the quality manager or the laboratory director) and the approval of the scope (the laboratory director).
On this copy each step on this page can be tried as one of the invented members of staff (the sample laboratory scientist, the sample quality manager, the sample laboratory director). The answer comes from the same checks and the same scoring as on a laboratory's own VERA. What is tried is held in the page, sent with each step, and gone on leaving it.
On a laboratory's own VERA the approval is also checked against the signed staff register.
Who takes each step, and why nobody else may
- Recording the evidence
A quality manager, laboratory scientist or laboratory director, each signed in.
Each record is one test file run by one operator on one reagent lot at one site. The same record entered twice is refused, and it is scored once.
- The evidence decision
The quality manager or laboratory director, signed in.
Nobody types the decision in: VERA scores the records under the evidence policy and records the score with its reason. Any other sign-in is refused, and so is a request with no sign-in.
- Approving the scope
The laboratory director only, signed in and listed in the signed staff register, with a licence number, as valid on the day.
The quality manager cannot approve: the decision is a score, and the approval is the laboratory director's own act. A laboratory director who recorded evidence of the campaign cannot approve it either: a person does not approve tests they performed.
- Withdrawing an approval
The laboratory director, signed in.
A withdrawal names the approval it ends, with a reason. Both stay on record; nothing is changed or deleted.