SU-000001
Sysmex UD-10
Read-only, byte-exact copy of the U-WAM ASTM result session; each result names the module that measured it (UD-10), so VERA attributes it to this asset. Images only through a Sysmex-documented export. VERA never answers a query or downloads an order.
ActiveTo set up on site

Release findings remain open.
The asset workbook records no room for this asset.
Extraction playbook
- Record the module serial, software version, U-WAM version, reagent and QC lots.
- Obtain the Sysmex U-WAM host-interface specification R315168 and freeze the message profile.
- Capture a read-only, byte-exact copy of the U-WAM ASTM result session (frames, checksums, ENQ/ACK/EOT); VERA never answers a query or downloads an order.
- Attribute each result to this module through R-14 and map host codes, units and grades through the signed Sysmex host-code table.
- Apply identity, QC, carryover, amended-result and flag-only gates before release.
Connection to VERA™
Awaiting connectionKERNEL is this analyser's host. Awaiting a channel on VERA's listener.
Instrument identity
- Manufacturer
- Sysmex
- Model
- UD-10
- Serial
- •••44
- Facility
- Unrecorded
- Evidence role
- analytical
- Software mode
- EXTERNAL VENDOR RUNTIME
Evidence contract
- Native evidence
- ASTM result records (RAW and MAINFORMAT), reviewed particle classifications, particle images when exportable, audit events
- Open derivatives
- ASTM E1394 host result stream (read-only copy), FHIR DiagnosticReport/Observation projection
- Vendor software
- UD-10 system software, Sysmex U-WAM (LIS2-A2, R315168)
- KG eligibility
- Disabled pending dual site activation
Acceptance tests
KERNEL simulator frames and the golden U-WAM fixtures yield every Check marker of this module with the right LOINC, unit and module.
AwaitedPhysical module inventory and serial match the register and the U-WAM configuration.
AwaitedQC and carryover failures block release.
AwaitedPrimary sources
Site activation attestations
No site approval is recorded in your session. Patient run creation remains disabled.
VERA refuses an approval of this instrument today, as it would at the lab:
- Its connection to VERA is not qualified yet: its connection is still to be set up on site.
- The Directory records no room for the UD-10.
- The version of its control software (U-WAM) is not recorded yet.
- Sysmex's specification for its connection is not yet in hand.
A revocation can still be recorded. Try an approval to see the rule bite.