SRA-SXA-000001
Sysmex haematology analyser
Read-only, byte-exact copy of the XR IPU's ASTM session with the host (KERNEL tee preferred, network tap as fallback); VERA never answers a query or downloads an order. Scattergram and research data only through a Sysmex-documented export.
ActiveTo set up on site

Extraction playbook
- Record the analyser serial, the IPU software version, the ASTM mode (not Sysmex Standard), reagent and QC lots.
- Obtain the Sysmex XR-Series host-interface specification R320120 and the extended-data export specification.
- Capture and seal the original ASTM sessions and any supported measurement/scattergram files.
- Map host codes to LOINC and UCUM through the signed Sysmex host-code table, keeping flags and comments verbatim.
- Apply identity, delta-check, QC, calibration and amended-result gates.
Connection to VERA™
Built, switched offVERA's listener is built and switched off for this analyser (ports 21003, 21004, 21009 and 21010): it reads a copy of the analyser's link once IT routes one to it on site. KERNEL stays its host; VERA only reads.
Instrument identity
- Manufacturer
- Sysmex
- Model
- XR-1000-1-A
- Serial
- •••03
- Facility
- Sample Reciving Area
- Evidence role
- analytical
- Software mode
- EXTERNAL VENDOR RUNTIME
Evidence contract
- Native evidence
- ASTM E1394 result records, IP messages and flags, scattergram/measurement channel outputs when exportable, QC, reagent lots, audit events
- Open derivatives
- ASTM E1394 host result stream (read-only copy), FHIR DiagnosticReport/Observation projection
- Vendor software
- XR-Series IPU software, Sysmex host interface (ASTM E1394-97, R320120)
- KG eligibility
- Disabled pending dual site activation
Acceptance tests
KERNEL simulator frames and the golden XR fixtures yield the 20 Check markers with the right LOINC and unit.
AwaitedCBC reference material and inter-instrument comparison meet the approved method contract; the differential is checked against manual counts (plan B2).
AwaitedNo result releases while required QC is invalid.
AwaitedPrimary sources
Site activation attestations
No site approval is recorded in your session. Patient run creation remains disabled.
VERA refuses an approval of this instrument today, as it would at the lab:
- Its connection to VERA is not qualified yet: its connection is still to be set up on site.
- The IPU software version is not recorded: read it from the IPU.
- The serial number is in the asset list of 17 Nov 2026.
- Its control computer must be set to send results in the standard laboratory format (ASTM), not Sysmex's own.
- Sysmex's specification for its connection is not yet in hand.
A revocation can still be recorded. Try an approval to see the rule bite.