SRA-SXA-000001

Sysmex haematology analyser

Read-only, byte-exact copy of the XR IPU's ASTM session with the host (KERNEL tee preferred, network tap as fallback); VERA never answers a query or downloads an order. Scattergram and research data only through a Sysmex-documented export.
ActiveTo set up on site
Laboratory room lined with enclosed instruments
Your sessionEmpty. What you run, approve or register on these pages: kept in this server's memory for 30 minutes after your last action, seen only by you, never written to disk.

Extraction playbook

  1. Record the analyser serial, the IPU software version, the ASTM mode (not Sysmex Standard), reagent and QC lots.
  2. Obtain the Sysmex XR-Series host-interface specification R320120 and the extended-data export specification.
  3. Capture and seal the original ASTM sessions and any supported measurement/scattergram files.
  4. Map host codes to LOINC and UCUM through the signed Sysmex host-code table, keeping flags and comments verbatim.
  5. Apply identity, delta-check, QC, calibration and amended-result gates.

Connection to VERA

Built, switched off

VERA's listener is built and switched off for this analyser (ports 21003, 21004, 21009 and 21010): it reads a copy of the analyser's link once IT routes one to it on site. KERNEL stays its host; VERA only reads.

Instrument identity

Manufacturer
Sysmex
Model
XR-1000-1-A
Serial
•••03
Facility
Sample Reciving Area
Evidence role
analytical
Software mode
EXTERNAL VENDOR RUNTIME

Evidence contract

Native evidence
ASTM E1394 result records, IP messages and flags, scattergram/measurement channel outputs when exportable, QC, reagent lots, audit events
Open derivatives
ASTM E1394 host result stream (read-only copy), FHIR DiagnosticReport/Observation projection
Vendor software
XR-Series IPU software, Sysmex host interface (ASTM E1394-97, R320120)
KG eligibility
Disabled pending dual site activation

Acceptance tests

KERNEL simulator frames and the golden XR fixtures yield the 20 Check markers with the right LOINC and unit.
Awaited
CBC reference material and inter-instrument comparison meet the approved method contract; the differential is checked against manual counts (plan B2).
Awaited
No result releases while required QC is invalid.
Awaited

Primary sources

Site activation attestations

No site approval is recorded in your session. Patient run creation remains disabled.
VERA refuses an approval of this instrument today, as it would at the lab:
  • Its connection to VERA is not qualified yet: its connection is still to be set up on site.
  • The IPU software version is not recorded: read it from the IPU.
  • The serial number is in the asset list of 17 Nov 2026.
  • Its control computer must be set to send results in the standard laboratory format (ASTM), not Sysmex's own.
  • Sysmex's specification for its connection is not yet in hand.
A revocation can still be recorded. Try an approval to see the rule bite.
A pseudonymous identity, such as qa.officer.1 or director.1: letters, digits and dots, never a name. The quality manager and the laboratory director must be two different people.

Demonstration copy