SRA-HBF-000001
Hamilton Blood Fractionation
Use the authenticated VENUS/LIM interface and preserve the exact method and run trace for every fractionation.
ActiveReady for the laboratory's validation

Extraction playbook
- Record input tube, output container, method checksum, operator, consumable lots, and VENUS version.
- Capture every barcode read, fraction destination, volume, liquid-level event, and intervention.
- Hash and seal the method and complete execution trace.
- Reconcile source tube to every derived aliquot in LIM.
- Block downstream evidence when identity, volume, or intervention policy fails.
Connection to VERA™
Reads its log filesVERA reads a Hamilton VENUS trace file from this machine when it is handed over as a file, and maps each line to a step with its sample and plate.
Instrument identity
- Manufacturer
- Hamilton
- Model
- Microlab STARlet EasyBlood
- Serial
- ••4k
- Facility
- Sample Reciving Area
- Evidence role
- pre-analytical
- Software mode
- EXTERNAL VENDOR RUNTIME
Evidence contract
- Native evidence
- method file, tube barcode events, fractionation run trace, volumes, errors, audit trail
- Open derivatives
- fraction lineage JSON, LIM process events
- Vendor software
- VENUS, EasyBlood method package
- KG eligibility
- Disabled pending dual site activation
Acceptance tests
Tube-to-aliquot lineage is complete and one-to-one with physical labels.
AwaitedVolume exceptions and manual overrides force review.
AwaitedThe approved method fixture reproduces the expected event sequence.
AwaitedPrimary sources
Site activation attestations
No site approval is recorded in your session. Patient run creation remains disabled.
VERA refuses an approval of this instrument today, as it would at the lab:
- It is not an analyser, so it cannot start a patient run.