SR-PR-000002
PacBio Revio
Use the supported Revio/SMRT Link transfer target and preserve the complete primary output bundle before secondary analysis.
ActiveReady for the laboratory's validation

Extraction playbook
- Record SMRT Cell, reagent lot, sample accession, run design, and exact software versions.
- Transfer every primary BAM, index, metadata, metric, and log file to quarantine.
- Create and seal the native bundle manifest before any alignment or variant calling.
- Generate open derivatives with a digest-pinned validated toolchain.
- Reconcile movie/run identifiers and read statistics against the primary report.
Connection to VERA™
Client built · awaiting connectionVERA's SMRT Link client (read-only: runs and SMRT Cells) and its reader for the share SMRT Link sends each cell to are built. SMRT Link and the share are to be set up on site.
Instrument identity
- Manufacturer
- PacBIO
- Model
- Revio
- Serial
- •••09
- Facility
- Sequencing Room
- Evidence role
- analytical
- Software mode
- EXTERNAL VENDOR RUNTIME
Evidence contract
- Native evidence
- HiFi reads BAM, PBI index, run metadata, primary analysis metrics, instrument logs
- Open derivatives
- FASTQ, CRAM, QC JSON
- Vendor software
- Revio system software, SMRT Link
- KG eligibility
- Disabled pending dual site activation
Acceptance tests
BAM quick-check, read-group, and PBI integrity checks pass.
AwaitedRun identifiers agree across BAM headers, metrics, and LIM accession.
AwaitedKnown truth-set material meets the approved HiFi precision and recall contract.
AwaitedPrimary sources
Site activation attestations
No site approval is recorded in your session. Patient run creation remains disabled.