SR-PR-000002

PacBio Revio

Use the supported Revio/SMRT Link transfer target and preserve the complete primary output bundle before secondary analysis.
ActiveReady for the laboratory's validation
Laboratory room lined with enclosed instruments
Your sessionEmpty. What you run, approve or register on these pages: kept in this server's memory for 30 minutes after your last action, seen only by you, never written to disk.

Extraction playbook

  1. Record SMRT Cell, reagent lot, sample accession, run design, and exact software versions.
  2. Transfer every primary BAM, index, metadata, metric, and log file to quarantine.
  3. Create and seal the native bundle manifest before any alignment or variant calling.
  4. Generate open derivatives with a digest-pinned validated toolchain.
  5. Reconcile movie/run identifiers and read statistics against the primary report.

Connection to VERA

Client built · awaiting connection

VERA's SMRT Link client (read-only: runs and SMRT Cells) and its reader for the share SMRT Link sends each cell to are built. SMRT Link and the share are to be set up on site.

Sequencer runs

Instrument identity

Manufacturer
PacBIO
Model
Revio
Serial
•••09
Facility
Sequencing Room
Evidence role
analytical
Software mode
EXTERNAL VENDOR RUNTIME

Evidence contract

Native evidence
HiFi reads BAM, PBI index, run metadata, primary analysis metrics, instrument logs
Open derivatives
FASTQ, CRAM, QC JSON
Vendor software
Revio system software, SMRT Link
KG eligibility
Disabled pending dual site activation

Acceptance tests

BAM quick-check, read-group, and PBI integrity checks pass.
Awaited
Run identifiers agree across BAM headers, metrics, and LIM accession.
Awaited
Known truth-set material meets the approved HiFi precision and recall contract.
Awaited

Primary sources

Site activation attestations

No site approval is recorded in your session. Patient run creation remains disabled.
A pseudonymous identity, such as qa.officer.1 or director.1: letters, digits and dots, never a name. The quality manager and the laboratory director must be two different people.

Demonstration copy