Genetic Sequencer (DNBSEQ-T20×2RS)

Copy the complete completed-run directory, including CAL, before ZTRON conversion or cleanup.
ActiveReady for the laboratory's validation
Laboratory room lined with enclosed instruments
Your sessionEmpty. What you run, approve or register on these pages: kept in this server's memory for 30 minutes after your last action, seen only by you, never written to disk.

Extraction playbook

  1. Record flow-cell, reagent lot, operator, instrument serial, and exact control/basecaller versions.
  2. Copy the complete run directory to quarantine without altering timestamps.
  3. Hash and seal CAL files, run configuration, cycle metrics, and logs.
  4. Run the validated ZTRON conversion package by digest to create FASTQ.
  5. Reconcile sample indices, read counts, and cycle metrics before release.

Connection to VERA

Reader built · awaiting connection

VERA's run-folder reader is built: it reads the DNBSEQ-T20's run folders on a read-only share and copies each run once every lane holds all its cycles and the folder has been quiet for 30 minutes. The share is to be set up on site.

Sequencer runs

Instrument identity

Manufacturer
MGI Tech Co., Ltd.
Model
DNBSEQ-T20×2RS
Serial
••••••••04
Facility
Sequencing Room
Evidence role
analytical
Software mode
EXTERNAL VENDOR RUNTIME

Evidence contract

Native evidence
CAL basecall binaries, run configuration, cycle metrics, instrument logs
Open derivatives
FASTQ, QC metrics
Vendor software
DNBSEQ-T20 control software, MGI basecalling, ZTRON
KG eligibility
Disabled pending dual site activation

Acceptance tests

CAL files exist for every expected lane and cycle.
Awaited
Sample-sheet identity is consistent with LIM accession and tube events.
Awaited
A fixed CAL fixture produces the approved FASTQ checksum set.
Awaited

Primary sources

Site activation attestations

No site approval is recorded in your session. Patient run creation remains disabled.
A pseudonymous identity, such as qa.officer.1 or director.1: letters, digits and dots, never a name. The quality manager and the laboratory director must be two different people.