SR-ASRS-000001
AVITI24 (Benchtop Short-Read Sequencer)
Read-only storage connection from the completed AVITI run folder; preserve the complete folder before conversion.
ActiveReady for the laboratory's validation

Extraction playbook
- Confirm run completion and the AVITI OS version in the run record.
- Transfer the complete run folder to the quarantine landing zone using a read-only storage connection.
- Calculate SHA-256 for every Bases file, AvitiRunStats.json, metadata file, and log before conversion.
- Seal the native folder manifest in immutable object storage.
- Run a validated Bases2Fastq image by digest to produce FASTQ and compare read counts against AVITI run statistics.
Connection to VERA™
Reader built · awaiting connectionVERA's run-folder reader is built: it reads the AVITI24's run folders on a read-only share and copies each run once RunUploaded.json, its last file, is written. The share is to be set up on site.
Instrument identity
- Manufacturer
- Element Biosciences, Inc
- Model
- 710-00975
- Serial
- ••••••03
- Facility
- Sequencing Room
- Evidence role
- analytical
- Software mode
- EXTERNAL VENDOR RUNTIME
Evidence contract
- Native evidence
- Bases files, AvitiRunStats.json, run parameters, instrument logs
- Open derivatives
- FASTQ, run QC JSON
- Vendor software
- AVITI OS, Bases2Fastq
- KG eligibility
- Disabled pending dual site activation
Acceptance tests
All expected lanes and samples are represented in the sealed manifest.
AwaitedFASTQ read counts reconcile with AvitiRunStats.json within the approved tolerance.
AwaitedA repeated conversion from the same Bases inputs produces identical declared outputs.
AwaitedPrimary sources
Site activation attestations
No site approval is recorded in your session. Patient run creation remains disabled.