SR-ASRS-000001

AVITI24 (Benchtop Short-Read Sequencer)

Read-only storage connection from the completed AVITI run folder; preserve the complete folder before conversion.
ActiveReady for the laboratory's validation
Laboratory room lined with enclosed instruments
Your sessionEmpty. What you run, approve or register on these pages: kept in this server's memory for 30 minutes after your last action, seen only by you, never written to disk.

Extraction playbook

  1. Confirm run completion and the AVITI OS version in the run record.
  2. Transfer the complete run folder to the quarantine landing zone using a read-only storage connection.
  3. Calculate SHA-256 for every Bases file, AvitiRunStats.json, metadata file, and log before conversion.
  4. Seal the native folder manifest in immutable object storage.
  5. Run a validated Bases2Fastq image by digest to produce FASTQ and compare read counts against AVITI run statistics.

Connection to VERA

Reader built · awaiting connection

VERA's run-folder reader is built: it reads the AVITI24's run folders on a read-only share and copies each run once RunUploaded.json, its last file, is written. The share is to be set up on site.

Sequencer runs

Instrument identity

Manufacturer
Element Biosciences, Inc
Model
710-00975
Serial
••••••03
Facility
Sequencing Room
Evidence role
analytical
Software mode
EXTERNAL VENDOR RUNTIME

Evidence contract

Native evidence
Bases files, AvitiRunStats.json, run parameters, instrument logs
Open derivatives
FASTQ, run QC JSON
Vendor software
AVITI OS, Bases2Fastq
KG eligibility
Disabled pending dual site activation

Acceptance tests

All expected lanes and samples are represented in the sealed manifest.
Awaited
FASTQ read counts reconcile with AvitiRunStats.json within the approved tolerance.
Awaited
A repeated conversion from the same Bases inputs produces identical declared outputs.
Awaited

Primary sources

Site activation attestations

No site approval is recorded in your session. Patient run creation remains disabled.
A pseudonymous identity, such as qa.officer.1 or director.1: letters, digits and dots, never a name. The quality manager and the laboratory director must be two different people.

Demonstration copy