LPR-NE-000001
NanoDrop Eight
Preserve the N8DB database and export full spectra plus unrounded measurements for each plate/sample map.
ActiveReady for the laboratory's validation

Extraction playbook
- Record instrument serial, software version, method, blank, pathlength, and plate/sample map.
- Copy the N8DB file after the run is closed and hash it in quarantine.
- Export measurements and all available spectrum data without spreadsheet reformatting.
- Reconcile well/sample identifiers with the LIM plate map.
- Apply concentration and purity QC gates without overwriting original values.
Connection to VERA™
PlannedPlanned: VERA reads its .tsv or .csv exports of DNA and RNA concentration and purity from a folder.
Instrument identity
- Manufacturer
- Thermo Fisher Scientific
- Model
- NanoDrop Eight
- Serial
- ••••••••50
- Facility
- Library Preparation Room
- Evidence role
- pre-analytical
- Software mode
- EXTERNAL VENDOR RUNTIME
Evidence contract
- Native evidence
- N8DB database with spectra and measurements, method, instrument audit/log data
- Open derivatives
- CSV/TSV measurements, spectrum exports
- Vendor software
- NanoDrop Eight PC software
- KG eligibility
- Disabled pending dual site activation
Acceptance tests
N8DB opens in the recorded software version and contains every expected well.
AwaitedOpen exports reproduce the native displayed values at full available precision.
AwaitedReference material and blank checks meet the approved method limits.
AwaitedPrimary sources
Site activation attestations
No site approval is recorded in your session. Patient run creation remains disabled.
VERA refuses an approval of this instrument today, as it would at the lab:
- It is not an analyser, so it cannot start a patient run.
A revocation can still be recorded. Try an approval to see the rule bite.