LPR-NE-000001

NanoDrop Eight

Preserve the N8DB database and export full spectra plus unrounded measurements for each plate/sample map.
ActiveReady for the laboratory's validation
Laboratory room lined with enclosed instruments
Your sessionEmpty. What you run, approve or register on these pages: kept in this server's memory for 30 minutes after your last action, seen only by you, never written to disk.

Extraction playbook

  1. Record instrument serial, software version, method, blank, pathlength, and plate/sample map.
  2. Copy the N8DB file after the run is closed and hash it in quarantine.
  3. Export measurements and all available spectrum data without spreadsheet reformatting.
  4. Reconcile well/sample identifiers with the LIM plate map.
  5. Apply concentration and purity QC gates without overwriting original values.

Connection to VERA

Planned

Planned: VERA reads its .tsv or .csv exports of DNA and RNA concentration and purity from a folder.

Instrument identity

Manufacturer
Thermo Fisher Scientific
Model
NanoDrop Eight
Serial
••••••••50
Facility
Library Preparation Room
Evidence role
pre-analytical
Software mode
EXTERNAL VENDOR RUNTIME

Evidence contract

Native evidence
N8DB database with spectra and measurements, method, instrument audit/log data
Open derivatives
CSV/TSV measurements, spectrum exports
Vendor software
NanoDrop Eight PC software
KG eligibility
Disabled pending dual site activation

Acceptance tests

N8DB opens in the recorded software version and contains every expected well.
Awaited
Open exports reproduce the native displayed values at full available precision.
Awaited
Reference material and blank checks meet the approved method limits.
Awaited

Primary sources

Site activation attestations

No site approval is recorded in your session. Patient run creation remains disabled.
VERA refuses an approval of this instrument today, as it would at the lab:
  • It is not an analyser, so it cannot start a patient run.
A revocation can still be recorded. Try an approval to see the rule bite.
A pseudonymous identity, such as qa.officer.1 or director.1: letters, digits and dots, never a name. The quality manager and the laboratory director must be two different people.

Demonstration copy