LPR-M-000003
MGISP-960
Use a supported authenticated interface or signed export and retain the complete execution trace before deriving sample lineage.
ActiveVendor interface needed

Extraction playbook
- Record asset, control software, firmware, method checksum, operator, deck/layout, and consumable lots.
- Capture worklist identity, barcode events, executed actions, errors, interventions, and timestamps.
- Hash and seal method and execution records before interpretation.
- Reconcile input sample identity to every output container or plate position.
- Block downstream evidence when trace, identity, method, or process-control checks fail.
Connection to VERA™
Reads its log filesVERA reads an MGI liquid-handler log from this machine when it is handed over as a file, and maps each line to a step with its sample and plate.
Instrument identity
- Manufacturer
- MGI Tech
- Model
- MGISP-960RS (version 5)
- Serial
- ••••••••28
- Facility
- Library Preparation Room
- Evidence role
- pre-analytical
- Software mode
- EXTERNAL VENDOR RUNTIME
Evidence contract
- Native evidence
- method/worklist, barcode events, executed steps, volumes or positions, errors, audit log
- Open derivatives
- sample lineage JSON, process-event JSON
- Vendor software
- Installed control software and method packages must be inventoried
- KG eligibility
- Disabled pending dual site activation
Acceptance tests
Positive, negative, duplicate, and mismatch fixtures produce the approved lineage states.
AwaitedManual interventions and errors remain visible and force the configured review.
AwaitedMethod or software changes invalidate the current validation decision.
AwaitedSite activation attestations
No site approval is recorded in your session. Patient run creation remains disabled.
VERA refuses an approval of this instrument today, as it would at the lab:
- Its connection to VERA is not qualified yet: the maker's interface package for its connection is not yet in hand.
- Still to do at the lab: obtain the vendor interface or export specification and exact installed software, firmware, and method versions.
- It is not an analyser, so it cannot start a patient run.
A revocation can still be recorded. Try an approval to see the rule bite.