CP-SU-000001

Sysmex UC-3500

Read-only, byte-exact copy of the U-WAM ASTM result session; each result names the module that measured it (UC-3500), so VERA attributes it to this asset. VERA never answers a query or downloads an order.
ActiveTo set up on site
Laboratory room lined with enclosed instruments
Your sessionEmpty. What you run, approve or register on these pages: kept in this server's memory for 30 minutes after your last action, seen only by you, never written to disk.

Extraction playbook

  1. Record the module serial, software version, U-WAM version, reagent and QC lots.
  2. Obtain the Sysmex U-WAM host-interface specification R315168 and freeze the message profile.
  3. Capture a read-only, byte-exact copy of the U-WAM ASTM result session (frames, checksums, ENQ/ACK/EOT); VERA never answers a query or downloads an order.
  4. Attribute each result to this module through R-14 and map host codes, units and grades through the signed Sysmex host-code table.
  5. Apply identity, QC, carryover, amended-result and flag-only gates before release.

Connection to VERA

Built, switched off

VERA's listener is built and switched off for this analyser (ports 21005, 21006, 21007 and 21008): it reads a copy of the analyser's link once IT routes one to it on site. KERNEL stays its host; VERA only reads.

Instrument identity

Manufacturer
Sysmex
Model
UC-3500
Serial
•••85
Facility
Clinical Pathology
Evidence role
analytical
Software mode
EXTERNAL VENDOR RUNTIME

Evidence contract

Native evidence
ASTM result records (RAW and MAINFORMAT), strip reflectance results, refractometry specific gravity, colour and turbidity, QC, reagent lots
Open derivatives
ASTM E1394 host result stream (read-only copy), FHIR DiagnosticReport/Observation projection
Vendor software
UC-3500 system software, Sysmex U-WAM (LIS2-A2, R315168)
KG eligibility
Disabled pending dual site activation

Acceptance tests

KERNEL simulator frames and the golden U-WAM fixtures yield every Check marker of this module with the right LOINC, unit and module.
Awaited
Physical module inventory and serial match the register and the U-WAM configuration.
Awaited
QC and carryover failures block release.
Awaited

Primary sources

Site activation attestations

No site approval is recorded in your session. Patient run creation remains disabled.
VERA refuses an approval of this instrument today, as it would at the lab:
  • Its connection to VERA is not qualified yet: its connection is still to be set up on site.
  • The version of its control software (U-WAM) is not recorded yet.
  • Sysmex's specification for its connection is not yet in hand.
A revocation can still be recorded. Try an approval to see the rule bite.
A pseudonymous identity, such as qa.officer.1 or director.1: letters, digits and dots, never a name. The quality manager and the laboratory director must be two different people.

Demonstration copy