CP-CP-000001
cobas pure
Receive vendor-supported host messages through a mutually authenticated interface and retain the original byte stream plus acknowledgements.
ActiveVendor interface needed

Extraction playbook
- Obtain Roche's interface specification for the exact installed software release and freeze the message profile.
- Capture original inbound and outbound messages with timestamps, message-control IDs, acknowledgements, and connection identity.
- Hash and seal each message envelope before terminology mapping.
- Map local test codes, units, flags, methods, and reference intervals under version control.
- Compare every normalised result to the source message and block duplicates, unit ambiguity, and missing specimen identity.
Connection to VERA™
Built, switched offVERA's listener is built and switched off for this analyser (ports 21001 and 21002): it reads a copy of the analyser's link once IT routes one to it on site. KERNEL stays its host; VERA only reads.
Instrument identity
- Manufacturer
- ROCHE
- Model
- cobas e 402 and cobas c 303
- Serial
- (21)U5K2-09 | (21)25E6-06
- Facility
- Clinical Pathology
- Evidence role
- analytical
- Software mode
- EXTERNAL VENDOR RUNTIME
Evidence contract
- Native evidence
- result messages, reaction flags, calibration and QC events, reagent lots, maintenance and audit events
- Open derivatives
- HL7 v2 ORU messages, FHIR DiagnosticReport/Observation projection
- Vendor software
- cobas pure system software, Roche host interface or middleware
- KG eligibility
- Disabled pending dual site activation
Acceptance tests
Golden HL7 fixtures round-trip without numeric, unit, flag, or precision loss.
AwaitedDuplicate and amended-result scenarios produce the approved state transitions.
AwaitedCalibration/QC lockouts prevent patient-evidence release.
AwaitedPrimary sources
Site activation attestations
No site approval is recorded in your session. Patient run creation remains disabled.
VERA refuses an approval of this instrument today, as it would at the lab:
- Its connection to VERA is not qualified yet: the maker's interface package for its connection is not yet in hand.
- Roche's guide to its connection, its installed versions and the fields it sends are not yet in hand.
- Raw detector signals are not assumed to be available through the host interface.
A revocation can still be recorded. Try an approval to see the rule bite.