CP-CP-000001

cobas pure

Receive vendor-supported host messages through a mutually authenticated interface and retain the original byte stream plus acknowledgements.
ActiveVendor interface needed
Laboratory room lined with enclosed instruments
Your sessionEmpty. What you run, approve or register on these pages: kept in this server's memory for 30 minutes after your last action, seen only by you, never written to disk.

Extraction playbook

  1. Obtain Roche's interface specification for the exact installed software release and freeze the message profile.
  2. Capture original inbound and outbound messages with timestamps, message-control IDs, acknowledgements, and connection identity.
  3. Hash and seal each message envelope before terminology mapping.
  4. Map local test codes, units, flags, methods, and reference intervals under version control.
  5. Compare every normalised result to the source message and block duplicates, unit ambiguity, and missing specimen identity.

Connection to VERA

Built, switched off

VERA's listener is built and switched off for this analyser (ports 21001 and 21002): it reads a copy of the analyser's link once IT routes one to it on site. KERNEL stays its host; VERA only reads.

Instrument identity

Manufacturer
ROCHE
Model
cobas e 402 and cobas c 303
Serial
(21)U5K2-09 | (21)25E6-06
Facility
Clinical Pathology
Evidence role
analytical
Software mode
EXTERNAL VENDOR RUNTIME

Evidence contract

Native evidence
result messages, reaction flags, calibration and QC events, reagent lots, maintenance and audit events
Open derivatives
HL7 v2 ORU messages, FHIR DiagnosticReport/Observation projection
Vendor software
cobas pure system software, Roche host interface or middleware
KG eligibility
Disabled pending dual site activation

Acceptance tests

Golden HL7 fixtures round-trip without numeric, unit, flag, or precision loss.
Awaited
Duplicate and amended-result scenarios produce the approved state transitions.
Awaited
Calibration/QC lockouts prevent patient-evidence release.
Awaited

Primary sources

Site activation attestations

No site approval is recorded in your session. Patient run creation remains disabled.
VERA refuses an approval of this instrument today, as it would at the lab:
  • Its connection to VERA is not qualified yet: the maker's interface package for its connection is not yet in hand.
  • Roche's guide to its connection, its installed versions and the fields it sends are not yet in hand.
  • Raw detector signals are not assumed to be available through the host interface.
A revocation can still be recorded. Try an approval to see the rule bite.