BSCR-HF-000001
Haier freezer, −20 °C
Use a supported read-only API or export. Preserve source timezone, sensor identity, calibration, alarms, and missing-data intervals.
ActiveVendor interface needed

Extraction playbook
- Record controller, probe, serial, location, software/firmware, timezone, and calibration identity.
- Export readings, alarms, set-point changes, and missing-data intervals at the highest supported resolution.
- Hash and seal each source export before time normalisation.
- Convert timestamps to UTC while preserving source timestamps and offsets.
- Associate excursions with affected sample and reagent custody windows only through explicit time/location joins.
Connection to VERA™
Reads its readings filesVERA reads this freezer's readings when they are handed over as a file and checks them against −25 to −15 °C.
Instrument identity
- Manufacturer
- Haier
- Model
- DW-25L262
- Serial
- ••••••••12
- Facility
- BSL-3 Stem Cell Room
- Evidence role
- environmental
- Software mode
- EXTERNAL VENDOR RUNTIME
Evidence contract
- Native evidence
- timestamped sensor readings, alarms, set-point changes, calibration records
- Open derivatives
- signed environmental event JSON, CSV time series
- Vendor software
- Installed controller/logger software must be inventoried
- KG eligibility
- Disabled pending dual site activation
Acceptance tests
Clock drift, timezone, calibration, and missing-data checks pass.
AwaitedAlarm events and source readings reconcile for a controlled excursion fixture.
AwaitedEnvironmental data cannot silently change an analytical value; it creates a release gate or qualifier.
AwaitedSite activation attestations
No site approval is recorded in your session. Patient run creation remains disabled.
VERA refuses an approval of this instrument today, as it would at the lab:
- Its connection to VERA is not qualified yet: the maker's interface package for its connection is not yet in hand.
- Still to do at the lab: obtain the exact controller or logger interface and installed firmware or software inventory.
- It is not an analyser, so it cannot start a patient run.
A revocation can still be recorded. Try an approval to see the rule bite.