Documentation · Working document

Requirements register

VERA-REQ-001 to 160 as the Operating Bible v3.0 states them, and VERA's own requirements since, each in or out of launch scope with the reason. A draft.

About this file

Kind of file
JSON data
Status
Working document
Version
1.0.0
Size
105 KB
File
requirements.v1.json
Fingerprint (SHA-256)
374dd1b838a83f6539a99644de729b13d8ffdf7a0f0af6c85b02ec3c0cee9f30

This page is drawn from the file itself each time it opens, so it always shows what the download holds.

VERA-REQ-001 to 160 are quoted word for word from the Operating Bible v3.0, and test names and file names as the code writes them, so they keep their own American spelling and terms such as “reference interval”.

About the register

Register
PREPAIRE-VERA-REQUIREMENTS@1.0.0, version 1.0.0
Review
Not yet signed. The laboratory director signs it, and the file then names who signed it and when
Owner and approver
The quality manager; the laboratory director
Requirements
196: 108 in launch scope (3 of them in part), 88 out
Source
Prepaire VERA Platform Operating Bible v3.0, Appendix A.1 Requirements register (sections A.1.1 to A.1.4), dated 24 August 2026, imported 15 November 2026

What the launch covers

VERA by Prepaire as the result path of GenetiQ Check at IRENA's clinical laboratory: blood and urine measured on the Roche cobas pure (c 303 and e 402) and the Sysmex XR-1000 and urine line (UC-3500, UF-5000, UD-10). VERA reads each analyser beside KERNEL and later hosts an analyser's link after the laboratory director's recorded takeover; it holds the order, the vial, QC, review, release per result, the lab report, the clinic message and the Malaffi message, the person's encrypted record and knowledge graph, and personal ranges, with the launch functions of VERA-REQ-161 onward. Robots, the world model, synthesis, autonomy, model results on patients, genetic results and the stool track are outside it.

Notes in the file

  • VERA's requirements register for the launch (F02 step 3; ISO 15189:2022 7.6.3 a)). VERA-REQ-001 to VERA-REQ-160 are imported word for word from Appendix A.1 of the Prepaire VERA Platform Operating Bible v3.0 (24 August 2026). VERA-REQ-161 onward are the launch functions built since. src/lib/validation/requirements.ts reads this file; docs/validation/master-plan.md explains the scope.
  • launchScope: in (a launch function: config/validation/trace.v1.json names the tests that prove it, and src/lib/validation/trace-matrix.test.ts fails when one has none) or out (with the reason). records: what a person must sign beyond the tests; config/validation/records.v1.json links each signed record.
  • The scope decisions are proposals for the quality manager and the laboratory director. The file ships pending; the laboratory director signs it in a reviewed change (reviewStatus signed_off, reviewer as a governed id, signedOn). Nothing here is signed for anyone.
  • Priorities P0 and P1 of VERA-REQ-121 to 160 are copied from the Requirements sheet of the Governance and Decision Register workbook; that sheet gives none for VERA-REQ-001 to 120 (null here). [to confirm] by the quality manager.

Requirements

196 requirements

The requirements register
RequirementWhat it saysLaunch scopePriorityTests and records
VERA-REQ-001AuthorityVera shall preserve LIMS authority for order, accession, patient-to-specimen, container, protocol, cancellation and quarantine.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeAt launch KERNEL is recommended to keep the accession, the patient's identity and cancellation (founder decision 5, a recommendation due 23 November 2026 [to confirm]); VERA also holds orders of its own (VERA-REQ-169). VERA never lets a LIMS cancellation or quarantine move a run it has released, and cancels an order only with a recorded reason.Not given
VERA-REQ-002IdentityVera shall use pseudonymous subject references and shall not copy unnecessary direct patient identifiers.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeEvery record of the result path names the person by a pseudonymous token; the identity vault holds the document number sealed.Not given
VERA-REQ-003CustodyEvery critical specimen transition shall bind expected and observed container, location, executor, station and time.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scope in partVERA records the vial's custody: who collected and received it, when, the container type, and the identity check at the draw.In scope: Who collected and who received the vial, when, its container type, and the identity check at the draw. VERA records no location or station of a transition: the collection SOP sets where (REC-003-1).Not givenRecord to sign: REC-003-1: SOP for specimen collection, transport and reception at IRENA (laboratory director)
VERA-REQ-004CustodyCritical mismatches shall produce no-go or quarantine; silent substitution is prohibited.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeA run whose sample, subject or specimen type does not match its vial holds every result (no-go); nothing is substituted.Not given
VERA-REQ-005LabelsLab-created specimen labels shall contain no PHI and shall use an approved machine-readable identifier profile.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceOut of launch scopeAt launch KERNEL is recommended to issue the accession numbers and tube labels (founder decision 5 of the go-live plan, a recommendation due 23 November 2026 [to confirm]); VERA prints no label and records no label event. [to confirm] when VERA issues its own labels.Not givenNone: out of launch scope
VERA-REQ-006ReagentsRun authorization shall verify reagent identity, lot, expiry, recall, quantity and position.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceOut of launch scopeReagent identity, lot, expiry and consumption are checked by the analysers and the laboratory's own procedures at launch; VERA records the QC material lot with each QC result. [to confirm] by the laboratory director.Not givenNone: out of launch scope
VERA-REQ-007AuthorizationEvery governed start shall require a signed, nonce-bound, expiring authorization envelope.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopePipeline steps run only under a workload token from the execution identity service: signed, for the execution audience, expiring.Not given
VERA-REQ-008AuthorizationAuthorization shall bind exact device, operator or robot, protocol, manifest, software and prerequisite state.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeA run starts only on an instrument with current dual activation for its exact assay protocol version, under an approved pipeline definition.Not given
VERA-REQ-009AuthorizationStale, replayed, reused or tampered authorizations shall fail.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeAn expired, malformed, wrongly addressed or foreign-signed token is refused.Not given
VERA-REQ-010SafetyVera shall use only vendor-supported start/worklist gates and shall not bypass functional safety.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeVERA only listens to the analysers until a takeover; as host it uses the vendor's own host interface (ASTM, HL7) and nothing else.Not givenRecord to sign: REC-010-1: Vendor host-interface manuals on file (Roche, Sysmex) (laboratory director)
VERA-REQ-011EdgeEdge adapters shall use unique device identity, least privilege, durable journal and signed updates.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeThe analyser listener is VERA's edge on the instrument network: one service identity, read-only, and a hash-chained message log.Not givenRecord to sign: REC-011-1: Image signatures recorded for each deployed image (CI sign-image job) (IT (release engineering))
VERA-REQ-012EdgeEdge capture shall continue safely through WAN loss and reconcile idempotently after recovery.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scope in partCapture runs on the node in the laboratory's own network, beside the analysers, so a loss of the wide-area link does not reach it [to confirm with IT: the analysers, the copy ports and the node share the laboratory network]. What VERA must show is reconciliation after an outage: a replay adds nothing twice.In scope: Reconciliation after recovery: a second replay of the KERNEL copy or of a captured stream adds nothing. That capture continues through a wide-area loss is the network's property, recorded by IT, not a VERA function.Not given
VERA-REQ-013EvidenceEvery required native artifact shall be hashed, manifest-linked, retained and recoverable.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeEvery artefact is hashed (SHA-256) and verified on completion, kept under the object store's lock and named in the release manifest.Not given
VERA-REQ-014EvidenceFile-close detection and post-transfer digest verification shall be validated per source.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scope in partAn upload completes only when its parts' digests and size verify; the sequencer watcher copies a run only once it is finished.In scope: Post-transfer digest verification of every upload: an upload completes only when its parts' digests and size verify. File-close detection of the sequencer watcher belongs to the stool plan (master plan 2.2), outside this one.Not given
VERA-REQ-015EvidenceCorrections shall be additive and preserve the superseded state.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeA correction is a new record: a corrected report revision keeps the original, and a correction supersedes a call-log entry with a reason.Not given
VERA-REQ-016EvidenceLarge biological artifacts shall remain off-chain and outside Knowledge Graph properties.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeThe knowledge graph holds released facts and pointers with checksums; raw files stay in the evidence store.Not given
VERA-REQ-017DataInbound HL7/ASTM source bytes shall be retained with message identity and acknowledgement state.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeThe listener keeps the exact bytes of every message with its direction, time and hash; the KERNEL copy keeps KERNEL's exact bytes.Not given
VERA-REQ-018DataNo observation shall be released without explicit unit, method, specimen, code provenance and source artifact.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeA clinical result is released only as a catalogue marker in its unit, LOINC-coded, from an artefact of the verified run.Not given
VERA-REQ-019DataMappings, conversions and reference intervals shall be versioned and reversible.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeEvery rule table carries its version, and every release records the digests of the tables that decided it.Not given
VERA-REQ-020DataMissing, not assessed, not detected, below limit, above limit and invalid shall remain distinct states.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeVERA keeps missing, not evaluable, below and above limit, and invalid as separate states, never as a number.Not given
VERA-REQ-021SciencePopulation interval, personal baseline and homeostatic target shall be separate data types.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeThe standard range, the person's own range and any target are separate values with separate labels.Not given
VERA-REQ-022ScienceAssociation, prediction, mechanism, intervention and causality shall be separate assertion types.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeFacts carry their assertion type; associations are never written as causal facts.Not given
VERA-REQ-023SoftwareProduction software shall be digest-pinned, signed, SBOM/provenance complete, licensed and validated.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeEvery pipeline build is pinned to its image digest and must pass its gates (signature, SBOM, licence, security review, rollback) before approval.Not givenRecord to sign: REC-023-1: SBOM, scan and signature of each VERA image (CI jobs vulnerability-scan and sign-image), kept with the release (IT (release engineering))
VERA-REQ-024SoftwareThe platform shall not use Docker-in-Docker or expose the host Docker socket to workloads.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeNo VERA service runs privileged or with the host's Docker socket; the sidecars run read-only without capabilities. One file is outside this requirement because it is never deployed: scripts/rehearsal/execution-node/compose.execution-node.yaml runs the execution node as one privileged container for a rehearsal on a Mac only (Docker on a Mac has no systemd to install the node on as the service of the host that it is on every deployed machine), and the kit's settings check refuses it on any other machine.Not given
VERA-REQ-025SoftwareVendor software execution mode and licensing rights shall be recorded explicitly.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeEach software package records its execution mode and licence; a model is used only under a recorded licence decision.Not given
VERA-REQ-026ModelsEvery model shall have immutable identity, intended use, training lineage, evaluation and rollback.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceOut of launch scopeNo model output is part of a launch result: AI explanations and model results stay off real patients until the DoH Responsible AI risk assessment (go-live plan D5).Not givenNone: out of launch scope
VERA-REQ-027ModelsModel output shall not independently authorize robot motion, synthesis, clinical release or treatment.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeA model result is a derived run that a person reviews and releases; it never authorises its own release.Not given
VERA-REQ-028ModelsModel and reference updates shall create new versions and impact analysis and shall not mutate history.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeA table or reference change makes a new version with a new digest; approved builds and recorded decisions never change.Not givenRecord to sign: REC-028-1: Change record of each table, model or reference change with its impact analysis (master plan section 7) (quality manager)
VERA-REQ-029ReleaseEvidence release shall be independently approved and signed.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeEvery release needs a qa officer or laboratory director who did not submit the run, and records who released it.Not given
VERA-REQ-030ReleaseKG release shall contain stable identifiers, typed assertions, provenance and object pointers.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeThe KG evidence bundle has stable ids, typed facts, provenance and object pointers, and is checked against its contract.Not given
VERA-REQ-031ReleaseRevoked or superseded evidence shall notify downstream systems and remain auditable.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeRevoking a release supersedes its graph write, cancels derived runs, retracts its messages and stays in the audit trail.Not given
VERA-REQ-032BioSafeSynthesis authorization shall verify user/customer legitimacy and approved sequence-screening policy.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceOut of launch scopeNo synthesis, printer, cartridge or BioSafe ledger takes part in the launch.Not givenNone: out of launch scope
VERA-REQ-033BioSafeBioSafe ledger payloads shall be minimized and exclude PHI, raw genomes and unnecessary sequence content.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceOut of launch scopeNo synthesis, printer, cartridge or BioSafe ledger takes part in the launch.Not givenNone: out of launch scope
VERA-REQ-034BioSafeA printer shall be enabled only through a supported interface after required endorsements.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceOut of launch scopeNo synthesis, printer, cartridge or BioSafe ledger takes part in the launch.Not givenNone: out of launch scope
VERA-REQ-035LedgerLedger membership, endorsement, revocation, backup and recovery governance shall be documented.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceOut of launch scopeNo BioSafe ledger takes part in the launch.Not givenNone: out of launch scope
VERA-REQ-036SecurityOT zones and conduits shall be default-deny and documented per device and robot.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeVERA's outbound connections stay on the node, the site and the signed UAE allow-list; the instrument network segment is IT's.Not givenRecord to sign: REC-036-1: Instrument network segment record: default-deny, per device (go-live A14) (IT)
VERA-REQ-037SecurityPeople, services, devices and robots shall use separate identity classes and credential lifecycles.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopePeople sign in through the identity service with roles; services and the analyser capture identity have their own identities and cannot act as people.Not given
VERA-REQ-038SecurityGenomic, metagenomic, operational imagery and detailed facility data shall receive restricted access and leakage monitoring.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceOut of launch scopeNo genomic, metagenomic or operational imaging data is part of the launch: genetic results follow genome-law consent (go-live plan D8), and stool runs as its own track (section 3.3, S1 to S8).Not givenNone: out of launch scope
VERA-REQ-039SecurityCryptographic receipts shall record algorithm, parameters, key ID and verification state.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeRelease receipts, report seals and graph seals record their algorithm and key id and are verified on reading.Not given
VERA-REQ-040SecurityPost-quantum migration shall be crypto-agile and evidence-based; transport claims shall not substitute for end-to-end controls.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceOut of launch scopeNo post-quantum migration is planned for the launch. Receipts and seals record their algorithm and key id (VERA-REQ-039), which keeps a later migration possible. [to confirm] by the information security officer.Not givenNone: out of launch scope
VERA-REQ-041QualityEvery lane, adapter, workflow, mapping, model and robot skill shall have a defined intended use and validation state.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeEach lane, adapter and table states what it is for and its status (pending, signed, demonstration); nothing unsigned decides a real result.Not given
VERA-REQ-042QualityMock adapters, simulated devices and timers shall be excluded from production or cryptographically isolated and labeled.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeSimulators, sample data and the demonstration sign-off are labelled, and every gate refuses them for a real subject.Not given
VERA-REQ-043QualityIQ/OQ/PQ, safety and negative-test evidence shall precede patient activation.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeOn an operational copy no clinical result is approved, released or reported by a build or scope the laboratory has not validated (IQ, OQ, PQ).Not givenRecord to sign: REC-043-1: Signed IQ, OQ and PQ records for the scope (laboratory director)
VERA-REQ-044QualityEvery critical change shall undergo impact assessment, regression and approval.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeEvery change goes through the change rule of the master plan: a new validated release, additive migrations, and a new campaign when the scope changes.Not givenRecord to sign: REC-044-1: Change record per release (impact, regression run, approval) (quality manager)
VERA-REQ-045OperationsPlatform health shall expose queue, disk, clock, certificate, transfer and reconciliation state.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeHealth and readiness name the build, mode, keys, database, object store, heartbeats and backlogs.Not given
VERA-REQ-046OperationsBackup, restore, rollback, state replay and reconciliation shall be exercised and evidenced.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeBackups, restores and rollbacks are exercised: the operations switches close release after a restore until its checks pass, and the reconcilers compare evidence with its off-host copy.Not givenRecord to sign: REC-046-1: Restore drill and rollback drill records (IT)
VERA-REQ-047OperationsBreak-glass shall require controlled approval and quarantine resulting evidence, without limiting independent emergency-stop capability.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceOut of launch scopeVERA has no break-glass path: an emergency changes nothing in VERA's gates, and the operations switches close the result path instead (VERA-REQ-046). [to confirm] by the laboratory director.Not givenNone: out of launch scope
VERA-REQ-048MaterialsReagent consumption and variance shall be traceable to run and lot.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceOut of launch scopeReagent identity, lot, expiry and consumption are checked by the analysers and the laboratory's own procedures at launch; VERA records the QC material lot with each QC result. [to confirm] by the laboratory director.Not givenNone: out of launch scope
VERA-REQ-049BillingBilling shall subscribe to signed usage evidence and shall not change safety or clinical states.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceOut of launch scopeNo billing system reads VERA's evidence at launch. The GenetiQ Partners events carry counts and pseudonyms only and change no clinical state (VERA-REQ-189).Not givenNone: out of launch scope
VERA-REQ-050GovernanceNo capability shall be labeled production, patient-active, autonomy-eligible or lights-out without exact approved scope and gate evidence.Operating Bible v3.0, A.1.1 Core authority, evidence and scienceIn launch scopeNothing is marked production or patient-active without its signed scope: unsigned tables hold every result, and only a validated build in a validated scope releases.Not given
VERA-REQ-051World ModelEvery operationally relevant physical object shall have a stable world_entity_id independent of mutable location and pose.Operating Bible v3.0, A.1.2 World model and spatial intelligenceOut of launch scopeThe launch has no physical world model or spatial twin: no robot workspace, robot frame or tracked physical object takes part in the result path.Not givenNone: out of launch scope
VERA-REQ-052World ModelEvery physical-state claim shall include observed time, source identity, coordinate frame where applicable, confidence or uncertainty and evidence references.Operating Bible v3.0, A.1.2 World model and spatial intelligenceOut of launch scopeThe launch has no physical world model or spatial twin: no robot workspace, robot frame or tracked physical object takes part in the result path.Not givenNone: out of launch scope
VERA-REQ-053World ModelExpected state and observed state shall remain separate and shall be reconciled through an explicit decision.Operating Bible v3.0, A.1.2 World model and spatial intelligenceOut of launch scopeThe launch has no physical world model or spatial twin: no robot workspace, robot frame or tracked physical object takes part in the result path.Not givenNone: out of launch scope
VERA-REQ-054World ModelCurrent world state shall be derived from append-only events and visible correction or supersession relationships.Operating Bible v3.0, A.1.2 World model and spatial intelligenceOut of launch scopeThe launch has no physical world model or spatial twin: no robot workspace, robot frame or tracked physical object takes part in the result path.Not givenNone: out of launch scope
VERA-REQ-055World ModelState freshness limits shall be defined per entity, relation, task and safety significance.Operating Bible v3.0, A.1.2 World model and spatial intelligenceOut of launch scopeThe launch has no physical world model or spatial twin: no robot workspace, robot frame or tracked physical object takes part in the result path.Not givenNone: out of launch scope
VERA-REQ-056World ModelPlanning and authorization shall bind to a declared world-state snapshot or digest and maximum acceptable age.Operating Bible v3.0, A.1.2 World model and spatial intelligenceOut of launch scopeThe launch has no physical world model or spatial twin: no robot workspace, robot frame or tracked physical object takes part in the result path.Not givenNone: out of launch scope
VERA-REQ-057World ModelWorld-model write interfaces shall accept signed observations or governed events and shall prohibit direct unversioned state overwrite.Operating Bible v3.0, A.1.2 World model and spatial intelligenceOut of launch scopeThe launch has no physical world model or spatial twin: no robot workspace, robot frame or tracked physical object takes part in the result path.Not givenNone: out of launch scope
VERA-REQ-058World ModelReservations shall not be treated as proof of occupancy, identity or readiness.Operating Bible v3.0, A.1.2 World model and spatial intelligenceOut of launch scopeThe launch has no physical world model or spatial twin: no robot workspace, robot frame or tracked physical object takes part in the result path.Not givenNone: out of launch scope
VERA-REQ-059World ModelAmbiguous, conflicting or unknown safety-critical state shall produce no-go, pause or quarantine.Operating Bible v3.0, A.1.2 World model and spatial intelligenceOut of launch scopeThe launch has no physical world model or spatial twin: no robot workspace, robot frame or tracked physical object takes part in the result path.Not givenNone: out of launch scope
VERA-REQ-060World ModelHistorical world state shall be replayable for incident investigation, validation and simulation.Operating Bible v3.0, A.1.2 World model and spatial intelligenceOut of launch scopeThe launch has no physical world model or spatial twin: no robot workspace, robot frame or tracked physical object takes part in the result path.Not givenNone: out of launch scope
VERA-REQ-061SpatialThe IRENA facility, zones, stations, instruments and robot workspaces shall be represented by controlled versioned geometry.Operating Bible v3.0, A.1.2 World model and spatial intelligenceOut of launch scopeThe launch has no physical world model or spatial twin: no robot workspace, robot frame or tracked physical object takes part in the result path.Not givenNone: out of launch scope
VERA-REQ-062SpatialEvery production coordinate transform shall have parent frame, calibration method, uncertainty, validity interval and approval state.Operating Bible v3.0, A.1.2 World model and spatial intelligenceOut of launch scopeThe launch has no physical world model or spatial twin: no robot workspace, robot frame or tracked physical object takes part in the result path.Not givenNone: out of launch scope
VERA-REQ-063SpatialRobot base, tool-center-point, station, instrument and sensor frames shall be uniquely identified and traceable.Operating Bible v3.0, A.1.2 World model and spatial intelligenceOut of launch scopeThe launch has no physical world model or spatial twin: no robot workspace, robot frame or tracked physical object takes part in the result path.Not givenNone: out of launch scope
VERA-REQ-064SpatialGeometry or calibration change outside approved tolerance shall suspend affected skill eligibility pending impact assessment.Operating Bible v3.0, A.1.2 World model and spatial intelligenceOut of launch scopeThe launch has no physical world model or spatial twin: no robot workspace, robot frame or tracked physical object takes part in the result path.Not givenNone: out of launch scope
VERA-REQ-065SpatialThe system shall detect material deviation between installed fixed assets and the approved spatial twin.Operating Bible v3.0, A.1.2 World model and spatial intelligenceOut of launch scopeThe launch has no physical world model or spatial twin: no robot workspace, robot frame or tracked physical object takes part in the result path.Not givenNone: out of launch scope
VERA-REQ-066SpatialEvery autonomy release shall declare a bounded operational design domain including facility, materials, tasks, people, environment, robot and fallback.Operating Bible v3.0, A.1.2 World model and spatial intelligenceOut of launch scopeThe launch has no physical world model or spatial twin: no robot workspace, robot frame or tracked physical object takes part in the result path.Not givenNone: out of launch scope
VERA-REQ-067SpatialProduction and simulation spatial registries shall be separated while retaining traceable correspondence.Operating Bible v3.0, A.1.2 World model and spatial intelligenceOut of launch scopeThe launch has no physical world model or spatial twin: no robot workspace, robot frame or tracked physical object takes part in the result path.Not givenNone: out of launch scope
VERA-REQ-068SpatialReachability, collision and route results shall retain geometry revision, solver version, tolerances and receipt.Operating Bible v3.0, A.1.2 World model and spatial intelligenceOut of launch scopeThe launch has no physical world model or spatial twin: no robot workspace, robot frame or tracked physical object takes part in the result path.Not givenNone: out of launch scope
VERA-REQ-069RoboticsEvery robot, controller and tool shall have exact identity, configuration, certificate, maintenance, calibration and eligibility state.Operating Bible v3.0, A.1.3 Robotics, action grammar, planning and perceptionOut of launch scopeNo robot, robot skill, planner, teleoperation or perception model takes part in the launch. Liquid handlers and robots are read as process logs only, which changes no result.Not givenNone: out of launch scope
VERA-REQ-070RoboticsEvery production robot action shall use a registered versioned skill with declared intended use and operational design domain.Operating Bible v3.0, A.1.3 Robotics, action grammar, planning and perceptionOut of launch scopeNo robot, robot skill, planner, teleoperation or perception model takes part in the launch. Liquid handlers and robots are read as process logs only, which changes no result.Not givenNone: out of launch scope
VERA-REQ-071RoboticsEvery governed robot action shall require a signed, expiring, single-use authorization bound to executor, tool, object, source, destination, skill, state and evidence.Operating Bible v3.0, A.1.3 Robotics, action grammar, planning and perceptionOut of launch scopeNo robot, robot skill, planner, teleoperation or perception model takes part in the launch. Liquid handlers and robots are read as process logs only, which changes no result.Not givenNone: out of launch scope
VERA-REQ-072RoboticsRobot execution parameters shall be schema-bound and range-limited; arbitrary script or free-form motion shall be prohibited in production.Operating Bible v3.0, A.1.3 Robotics, action grammar, planning and perceptionOut of launch scopeNo robot, robot skill, planner, teleoperation or perception model takes part in the launch. Liquid handlers and robots are read as process logs only, which changes no result.Not givenNone: out of launch scope
VERA-REQ-073RoboticsRobot completion shall require declared postcondition verification and shall not rely solely on controller success.Operating Bible v3.0, A.1.3 Robotics, action grammar, planning and perceptionOut of launch scopeNo robot, robot skill, planner, teleoperation or perception model takes part in the launch. Liquid handlers and robots are read as process logs only, which changes no result.Not givenNone: out of launch scope
VERA-REQ-074RoboticsEvery critical physical action shall produce a signed action receipt linked to authorization, pre/post state and evidence.Operating Bible v3.0, A.1.3 Robotics, action grammar, planning and perceptionOut of launch scopeNo robot, robot skill, planner, teleoperation or perception model takes part in the launch. Liquid handlers and robots are read as process logs only, which changes no result.Not givenNone: out of launch scope
VERA-REQ-075RoboticsRobot and station controllers shall accept only allow-listed program identities or supported operation calls.Operating Bible v3.0, A.1.3 Robotics, action grammar, planning and perceptionOut of launch scopeNo robot, robot skill, planner, teleoperation or perception model takes part in the launch. Liquid handlers and robots are read as process logs only, which changes no result.Not givenNone: out of launch scope
VERA-REQ-076RoboticsRetries and recoveries shall be bounded, declared and separately evidenced; unbounded retry loops are prohibited.Operating Bible v3.0, A.1.3 Robotics, action grammar, planning and perceptionOut of launch scopeNo robot, robot skill, planner, teleoperation or perception model takes part in the launch. Liquid handlers and robots are read as process logs only, which changes no result.Not givenNone: out of launch scope
VERA-REQ-077RoboticsTeleoperation shall use a separate mode, trained identity, authorization, limits and complete recording.Operating Bible v3.0, A.1.3 Robotics, action grammar, planning and perceptionOut of launch scopeNo robot, robot skill, planner, teleoperation or perception model takes part in the launch. Liquid handlers and robots are read as process logs only, which changes no result.Not givenNone: out of launch scope
VERA-REQ-078RoboticsA robot shall not manipulate undocumented internal instrument mechanisms or bypass vendor-supported handoff boundaries.Operating Bible v3.0, A.1.3 Robotics, action grammar, planning and perceptionOut of launch scopeNo robot, robot skill, planner, teleoperation or perception model takes part in the launch. Liquid handlers and robots are read as process logs only, which changes no result.Not givenNone: out of launch scope
VERA-REQ-079RoboticsRobot, tool and station maintenance or calibration overdue state shall block affected action eligibility.Operating Bible v3.0, A.1.3 Robotics, action grammar, planning and perceptionOut of launch scopeNo robot, robot skill, planner, teleoperation or perception model takes part in the launch. Liquid handlers and robots are read as process logs only, which changes no result.Not givenNone: out of launch scope
VERA-REQ-080RoboticsA robot carrying biological material shall use approved routes, payload retention and safe-hold locations.Operating Bible v3.0, A.1.3 Robotics, action grammar, planning and perceptionOut of launch scopeNo robot, robot skill, planner, teleoperation or perception model takes part in the launch. Liquid handlers and robots are read as process logs only, which changes no result.Not givenNone: out of launch scope
VERA-REQ-081Action GrammarThe production action grammar shall be finite, versioned, typed and machine-readable.Operating Bible v3.0, A.1.3 Robotics, action grammar, planning and perceptionOut of launch scopeNo robot, robot skill, planner, teleoperation or perception model takes part in the launch. Liquid handlers and robots are read as process logs only, which changes no result.Not givenNone: out of launch scope
VERA-REQ-082Action GrammarEach primitive shall define inputs, preconditions, parameters, controller mapping, safety envelope, postconditions, evidence, timeouts and recovery.Operating Bible v3.0, A.1.3 Robotics, action grammar, planning and perceptionOut of launch scopeNo robot, robot skill, planner, teleoperation or perception model takes part in the launch. Liquid handlers and robots are read as process logs only, which changes no result.Not givenNone: out of launch scope
VERA-REQ-083Action GrammarThe workflow compiler shall emit only primitives eligible for the exact robot, tool, station, object and operational design domain.Operating Bible v3.0, A.1.3 Robotics, action grammar, planning and perceptionOut of launch scopeNo robot, robot skill, planner, teleoperation or perception model takes part in the launch. Liquid handlers and robots are read as process logs only, which changes no result.Not givenNone: out of launch scope
VERA-REQ-084Action GrammarIdentity-sensitive movement shall use an approved identity verification before and after the critical transfer.Operating Bible v3.0, A.1.3 Robotics, action grammar, planning and perceptionOut of launch scopeNo robot, robot skill, planner, teleoperation or perception model takes part in the launch. Liquid handlers and robots are read as process logs only, which changes no result.Not givenNone: out of launch scope
VERA-REQ-085PlanningA planner shall propose actions only and shall not possess direct actuation credentials.Operating Bible v3.0, A.1.3 Robotics, action grammar, planning and perceptionOut of launch scopeNo robot, robot skill, planner, teleoperation or perception model takes part in the launch. Liquid handlers and robots are read as process logs only, which changes no result.Not givenNone: out of launch scope
VERA-REQ-086PlanningReplanning shall remain within pre-approved alternatives and shall trigger reauthorization when action bindings or safety-relevant state change.Operating Bible v3.0, A.1.3 Robotics, action grammar, planning and perceptionOut of launch scopeNo robot, robot skill, planner, teleoperation or perception model takes part in the launch. Liquid handlers and robots are read as process logs only, which changes no result.Not givenNone: out of launch scope
VERA-REQ-087PlanningPlanner infeasibility and alternative selection shall be explainable and linked to state, constraint and policy evidence.Operating Bible v3.0, A.1.3 Robotics, action grammar, planning and perceptionOut of launch scopeNo robot, robot skill, planner, teleoperation or perception model takes part in the launch. Liquid handlers and robots are read as process logs only, which changes no result.Not givenNone: out of launch scope
VERA-REQ-088PerceptionEvery production perception output shall bind sensor, calibration, model/configuration digest, time, frame and evidence.Operating Bible v3.0, A.1.3 Robotics, action grammar, planning and perceptionOut of launch scopeNo robot, robot skill, planner, teleoperation or perception model takes part in the launch. Liquid handlers and robots are read as process logs only, which changes no result.Not givenNone: out of launch scope
VERA-REQ-089PerceptionCritical identity shall use approved independent machine-readable evidence unless lane validation explicitly supports another method.Operating Bible v3.0, A.1.3 Robotics, action grammar, planning and perceptionOut of launch scopeNo robot, robot skill, planner, teleoperation or perception model takes part in the launch. Liquid handlers and robots are read as process logs only, which changes no result.Not givenNone: out of launch scope
VERA-REQ-090PerceptionPose estimates used for action shall include uncertainty within skill-specific tolerance.Operating Bible v3.0, A.1.3 Robotics, action grammar, planning and perceptionOut of launch scopeNo robot, robot skill, planner, teleoperation or perception model takes part in the launch. Liquid handlers and robots are read as process logs only, which changes no result.Not givenNone: out of launch scope
VERA-REQ-091PerceptionPerception validation shall report performance by relevant object, material, lighting, occlusion, damage and out-of-distribution condition.Operating Bible v3.0, A.1.3 Robotics, action grammar, planning and perceptionOut of launch scopeNo robot, robot skill, planner, teleoperation or perception model takes part in the launch. Liquid handlers and robots are read as process logs only, which changes no result.Not givenNone: out of launch scope
VERA-REQ-092PerceptionProduction perception models shall not adapt online without approved dataset, validation, release and rollback.Operating Bible v3.0, A.1.3 Robotics, action grammar, planning and perceptionOut of launch scopeNo robot, robot skill, planner, teleoperation or perception model takes part in the launch. Liquid handlers and robots are read as process logs only, which changes no result.Not givenNone: out of launch scope
VERA-REQ-093PerceptionSensor disagreement and out-of-distribution state shall be explicit and shall route according to approved policy.Operating Bible v3.0, A.1.3 Robotics, action grammar, planning and perceptionOut of launch scopeNo robot, robot skill, planner, teleoperation or perception model takes part in the launch. Liquid handlers and robots are read as process logs only, which changes no result.Not givenNone: out of launch scope
VERA-REQ-094PerceptionOperational camera collection shall be purpose-scoped, minimized, access-controlled and retention-governed.Operating Bible v3.0, A.1.3 Robotics, action grammar, planning and perceptionOut of launch scopeNo operational camera collects images at launch.Not givenNone: out of launch scope
VERA-REQ-095SafetyRobot motion safety shall remain with validated local safety-rated controls and shall be independent of enterprise or public-cloud availability.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeRobot motion safety: no robot takes part in the launch.Not givenNone: out of launch scope
VERA-REQ-096SafetyEvery autonomous lane shall have a documented hazard analysis, safety requirements, verification evidence and accepted residual risk.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeNo autonomous lane or closed cell is part of the launch: every clinical result is reviewed and released by people.Not givenNone: out of launch scope
VERA-REQ-097SafetyProtective stop, controlled abort, emergency stop and facility emergency states shall have distinct behavior and restart authority.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeRobot motion safety: no robot takes part in the launch.Not givenNone: out of launch scope
VERA-REQ-098SafetyHuman entry, obstacle, unexpected contact, spill and containment states shall trigger the validated safe response.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeRobot motion safety: no robot takes part in the launch.Not givenNone: out of launch scope
VERA-REQ-099SafetyEmergency-stop capability shall not depend on Vera authorization, network connectivity or language-model output.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeRobot motion safety: no robot takes part in the launch.Not givenNone: out of launch scope
VERA-REQ-100SafetyRobots shall preserve approved sterile, containment, clean/dirty and waste route boundaries.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeRobot motion safety: no robot takes part in the launch.Not givenNone: out of launch scope
VERA-REQ-101SafetyA dropped, leaking, uncapped or unidentified specimen shall trigger protective response and trained human disposition.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeRobot motion safety: no robot takes part in the launch.Not givenNone: out of launch scope
VERA-REQ-102SafetyCollaborative or shared-workspace operation shall be task-specifically risk-assessed and validated for exact speed, force, geometry and human exposure.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeRobot motion safety: no robot takes part in the launch.Not givenNone: out of launch scope
VERA-REQ-103SimulationEvery production-relevant simulation episode shall record twin revision, scenario, seed, software/model identities, initial state, result and artifacts.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeRobot and cell simulation: no robot or simulated cell takes part in the launch.Not givenNone: out of launch scope
VERA-REQ-104SimulationSimulation fidelity, assumptions and known gaps shall be documented and shall not substitute for real hardware qualification.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeRobot and cell simulation: no robot or simulated cell takes part in the launch.Not givenNone: out of launch scope
VERA-REQ-105SimulationSoftware-in-the-loop and hardware-in-the-loop testing shall cover nominal, negative, outage, safety, security and recovery scenarios.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeRobot and cell simulation: no robot or simulated cell takes part in the launch.Not givenNone: out of launch scope
VERA-REQ-106SimulationSim-to-real discrepancy thresholds shall be defined before acceptance testing and shall be measured.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeRobot and cell simulation: no robot or simulated cell takes part in the launch.Not givenNone: out of launch scope
VERA-REQ-107SimulationSignificant real incidents and near misses shall create or update controlled replay scenarios and regression coverage.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeEvery past incident of the result path is a regression test: the August 2026 export (protein 1,248 g/L, calcium judged against the wrong unit) is quarantined.Not given
VERA-REQ-108LearningProduction-derived training cases shall undergo privacy, provenance, labeling, split and bias governance before model development.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeNo model is trained or promoted on production data at launch: research use and the learning loop follow ethics approval (go-live plan D6).Not givenNone: out of launch scope
VERA-REQ-109LearningModel or policy promotion shall pass independent evaluation, relevant simulation/HIL, controlled deployment and rollback approval.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeNo model is trained or promoted on production data at launch: research use and the learning loop follow ethics approval (go-live plan D6).Not givenNone: out of launch scope
VERA-REQ-110BioSafeBioSafe robot actions shall bind customer/task authorization to exact robot, tool, material, printer, cartridge and state.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeNo synthesis, printer, cartridge or BioSafe ledger takes part in the launch.Not givenNone: out of launch scope
VERA-REQ-111BioSafeA robot shall not load sequence-related material into a printer or synthesis station without valid task and material authorization.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeNo synthesis, printer, cartridge or BioSafe ledger takes part in the launch.Not givenNone: out of launch scope
VERA-REQ-112BioSafeBiosecurity no-go or revocation shall propagate to robot, material, printer and workflow state before fulfillment.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeNo synthesis, printer, cartridge or BioSafe ledger takes part in the launch.Not givenNone: out of launch scope
VERA-REQ-113OperationsThe operations console shall expose live world-state discrepancy, robot mode, tool, payload, task, safety, maintenance and evidence state.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeNo robot, robot skill, planner, teleoperation or perception model takes part in the launch. Liquid handlers and robots are read as process logs only, which changes no result.Not givenNone: out of launch scope
VERA-REQ-114OperationsUnattended patient-active operation shall have a 24x7 alert, response and accountable on-call path.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeA stopped process, a failed backup or a backlog raises an alert to a named person.Not givenRecord to sign: REC-114-1: Named on-call rota and one tabletop incident exercise (go-live B18) (executive go-live authority)
VERA-REQ-115OperationsEvery autonomous lane shall have validated manual or alternative continuity procedures for essential work.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeNo autonomous lane or closed cell is part of the launch: every clinical result is reviewed and released by people.Not givenNone: out of launch scope
VERA-REQ-116OperationsHuman intervention shall be measured, categorized and visible; hidden manual work shall not support an autonomy claim.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeNo autonomous lane or closed cell is part of the launch: every clinical result is reviewed and released by people.Not givenNone: out of launch scope
VERA-REQ-117AutonomyAutonomy level shall be assigned per exact lane and operational design domain and shall be reducible after degradation.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeNo autonomous lane or closed cell is part of the launch: every clinical result is reviewed and released by people.Not givenNone: out of launch scope
VERA-REQ-118AutonomyAdvancement to closed autonomous cell shall require demonstrated safe normal flow, prescribed exception handling and complete evidence.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeNo autonomous lane or closed cell is part of the launch: every clinical result is reviewed and released by people.Not givenNone: out of launch scope
VERA-REQ-119AutonomyThe first IRENA closed-cell release shall meet the approved consecutive-sample acceptance target with zero critical substitutions, unidentified transitions or unauthorized actions.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeNo autonomous lane or closed cell is part of the launch: every clinical result is reviewed and released by people.Not givenNone: out of launch scope
VERA-REQ-120GovernanceEvery go-live decision shall name exact site, lane, robot, controller, tool, station, instrument, materials, software, models, safety configuration, hours, autonomy level and limitations.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeEach scope's go-live decision names the analyser, assay protocol version, VERA image digests, tables and limitations: the validation campaign records them, and the release refuses another.Not givenRecord to sign: REC-120-1: Signed evidence pack and go-live decision per scope (go-live C6) (laboratory director)
VERA-REQ-121ProductVERA Core, VERA Edge, VERA S1, BioSafe Custody, LIM Ingestor and the KG/Twin Gateway shall have explicit non-overlapping authority and versioned interfaces.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeKERNEL, VERA, the partner port and the knowledge graph have separate authority; every interface is a versioned contract.P1
VERA-REQ-122EdgeEach production edge lane shall bind to one exact instrument asset, serial, module set, firmware, workstation software and validated adapter baseline.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeEach run binds one register entry (asset id and serial); the release refuses a bundle from another instrument.P0Record to sign: REC-122-1: Nameplate readings: serials and software versions of each analyser (go-live A7) (laboratory director)
VERA-REQ-123EdgeThe instrument-facing edge interface shall be read-only unless a separate vendor-supported, safety-reviewed and validated control contract is approved.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeThe listener only listens; VERA hosts an analyser's link only after the laboratory director's recorded takeover for that analyser.P0
VERA-REQ-124EdgeNo edge node shall create a routable path from an instrument control network to the public internet, enterprise user network or Knowledge Graph.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeThe listener accepts only its copy sources and drops an analyser's own address; the node's egress stays on the allow-list.P0Record to sign: REC-124-1: Network record: no route from the instrument network to the internet, the office network or the graph (IT)
VERA-REQ-125EdgeEvery paired node shall use separate instrument-facing and VERA-facing zones, unique device identity, least privilege and default-deny conduits.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeEach channel names its copy source; the listener and the analyser host bind to their own addresses.P0Record to sign: REC-125-1: Zone and conduit record for the VERA node on the instrument network (go-live A14) (IT)
VERA-REQ-126EdgeLocal spooling shall be encrypted, bounded, monitored, power-failure tolerant and reconciled by content hash after recovery.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeCaptured bytes are sealed under the capture key, hash-chained and reconciled by hash after a replay.P0
VERA-REQ-127HardwareDGX Spark use shall be gated by architecture, driver, accelerator, storage, network, thermal, UPS and vendor-software compatibility evidence.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeVERA runs on DGX Spark nodes; the IQ record names the node, its mode, image digests and configuration before validation.P1Record to sign: REC-127-1: Hardware record per Spark: architecture, driver, accelerator, storage, network, thermal limits, UPS, disk encryption, clock and the vendor software's compatibility (go-live A13, A14) (IT)
VERA-REQ-128HardwareA qualified companion runtime may execute unsupported x86_64 or Windows packages while preserving the same signed plan, isolation and receipt contract.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeNo vendor package runs on a companion or external runtime at launch: the blood and urine results come from the analysers' own messages.P1None: out of launch scope
VERA-REQ-129SoftwareVendor software shall not be containerized, copied or automated beyond licensed rights and written vendor support.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeA vendor package runs only in the mode its licence allows; report-only packages are blocked.P1
VERA-REQ-130SoftwareExternal vendor runtimes shall produce an immutable execution manifest with binary identity, version, host identity, inputs, configuration, outputs and verification state.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeNo vendor package runs on a companion or external runtime at launch: the blood and urine results come from the analysers' own messages.P1None: out of launch scope
VERA-REQ-131SoftwareEvery production package shall declare supported CPU architecture, operating system, accelerator, driver, license, network dependency and hardware binding.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeEvery execution profile declares its architecture, CPU, memory and GPU; the platform refuses an incomplete one.P1
VERA-REQ-132EvidenceEvery lane shall define a modality-specific required-artifact manifest covering native, open, derivative, QC, configuration, audit and transport objects.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeEach lane declares the outputs a release needs; a run without them cannot release.P0
VERA-REQ-133EvidenceVendor conclusions shall be retained as signed comparator evidence and shall not overwrite native or independent analytical evidence.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeKERNEL's results are kept as a separate copy and compared with VERA's own reading; neither overwrites the other.P0
VERA-REQ-134EvidenceNative artifacts shall be sealed before transformation; every derivative shall bind source checksum, recipe, software, model, database and reference snapshot.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeNative bytes are sealed first; each derived fact names the checksum of what it came from.P0
VERA-REQ-135EvidenceA clinical message or PDF report shall never be represented as measurement-level raw evidence when richer source artifacts exist or are contractually required.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeA PDF report brought in is kept as a historical import, never as a measurement of the lab.P0
VERA-REQ-136DataEvery dataset dependency shall have an approved snapshot ID, checksum, source URL, access tier, license, permitted use, population description and withdrawal plan.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeEach reference dataset VERA uses (biological variation, cohorts, the reference snapshot) names its source, version, digest and permission.P1
VERA-REQ-137DataFixture, truth, cohort, calibration, reference-database and KG-enrichment datasets shall remain distinct governance classes.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeSynthetic fixtures, reference data and a person's own results are separate classes; fixtures are marked synthetic and refused for a real subject.P1
VERA-REQ-138DataControlled-access or cloud-bound data shall not be copied into VERA unless the approved access agreement and technical environment permit it.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeNo data goes outside the node, the site and the signed UAE allow-list; analysis APIs keep native data and genomes at home.P1
VERA-REQ-139ScienceModel selection shall use a predeclared benchmark protocol, frozen truth sets, acceptance metrics, subgroup analysis, reproducibility and rollback criteria.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeNo model is selected for a launch result (see VERA-REQ-026). The benchmark protocol applies when a model is proposed for patients.P1None: out of launch scope
VERA-REQ-140SciencePopulation reference intervals, cohort-conditioned context, personal longitudinal baselines and homeostatic targets shall use separate schemas and clinical labels.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeThe standard range and the person's own range use separate fields and labels on every screen, report and graph.P1
VERA-REQ-141ScienceNo personal baseline shall be created until minimum longitudinal density, stability, analytical comparability and review criteria are met.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeA personal range is drawn only once the signed minimum of earlier eligible results exists; before that VERA says why it stays on the standard range.P1
VERA-REQ-142ScienceCross-axis outputs shall expose contributing evidence, direction, uncertainty, missingness, model state and assertion type.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeCross-axis multi-omic outputs are not part of the launch result path. The CKM and organ-age readings that combine a person's results are covered by VERA-REQ-183.P1None: out of launch scope
VERA-REQ-143ScienceCausal claims shall require a separately approved evidence rule and shall not be inferred from correlation or model attention alone.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeThe reference graph carries the assertion type of every link; VERA never writes a causal claim from a correlation.P1
VERA-REQ-144ModelsIncoming patient assays may update governed patient state but shall not directly update production model weights, thresholds or reference snapshots.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeA person's results change only that person's graph and ranges; no shared model, threshold or reference changes with them.P1
VERA-REQ-145ModelsTraining candidates derived from production shall pass consent, de-identification, provenance, labeling, split, bias and independent-evaluation controls.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeNo model is trained or promoted on production data at launch: research use and the learning loop follow ethics approval (go-live plan D6).P1None: out of launch scope
VERA-REQ-146ReleaseOnly clinical.released or scientifically approved evidence states may create patient-graph assertions; held or quarantined runs shall have an empty write set.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeOnly a released run writes to the person's graph; a held or quarantined run writes nothing.P0
VERA-REQ-147ReleaseLarge source and derivative objects shall remain in an encrypted evidence vault addressed by resolvable pointer and checksum.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeLarge objects stay in the encrypted evidence store, addressed by pointer and checksum.P0
VERA-REQ-148ReleaseEvery KG/Twin update shall bind source artifacts, execution receipt, normalization plan, reference snapshot, review identity and release signature.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeEach graph write binds the release, its bundle digest, the reviewer and the build.P0
VERA-REQ-149CustodyPersonal mobile phones shall not participate in the clinical custody chain; scanners and printers shall be managed, inventoried and remotely governable.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeAt launch KERNEL is recommended to issue the accession numbers and tube labels (founder decision 5 of the go-live plan, a recommendation due 23 November 2026 [to confirm]); VERA prints no label and records no label event. [to confirm] when VERA issues its own labels.P0None: out of launch scope
VERA-REQ-150CustodyA machine-readable scan shall assist but not replace required human-readable label confirmation and protocol-specific identity checks.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeThe person is identified at the draw against the identity vault; the barcode on the tube must be the vial's own.P0Record to sign: REC-150-1: SOP for identification at collection and label confirmation (laboratory director)
VERA-REQ-151CustodyEvery label issuance, application, void, reprint, damage and disposal event shall be reconciled and auditable.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeAt launch KERNEL is recommended to issue the accession numbers and tube labels (founder decision 5 of the go-live plan, a recommendation due 23 November 2026 [to confirm]); VERA prints no label and records no label event. [to confirm] when VERA issues its own labels.P0None: out of launch scope
VERA-REQ-152BioSafeBioSafe ledger records shall contain only minimal pseudonymous commitments and shall not be required for emergency stop or local safe-state enforcement.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeNo BioSafe ledger takes part in the launch.P1None: out of launch scope
VERA-REQ-153BioSafeLedger outage shall invoke a documented continuity mode with bounded local evidence capture, no unauthorized new starts and deterministic retrospective reconciliation.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsOut of launch scopeNo BioSafe ledger takes part in the launch.P1None: out of launch scope
VERA-REQ-154SafetyVERA shall never cut instrument mains power, alter safety circuits, patch firmware or interrupt active acquisition unless the vendor-supported interface and safety case explicitly permit it.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeVERA sends nothing to an analyser except, as host, the vendor's own order and acknowledgement messages: no power, firmware or safety command.P1
VERA-REQ-155ValidationEach edge lane shall pass native restore, deterministic rerun, vendor-versus-independent comparison, duplicate, negative, interruption, corruption, rollback and quarantine tests.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeEach analyser lane passes byte replay, duplicate, corrupt-frame, KERNEL-versus-VERA comparison and quarantine tests.P1
VERA-REQ-156ValidationThe first sequencing and clinical lanes shall meet approved acceptance criteria before a scale decision for the remaining five lanes.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeThe Roche chemistry scope goes first and must meet the acceptance limits before the next scope.P1Record to sign: REC-156-1: Signed evidence pack of the first scope (go-live C6) (laboratory director)
VERA-REQ-157ValidationNo lane shall be described as production until the release packet identifies exact configuration, validation evidence, limitations, owner, effective date and rollback path.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeA scope is production only with its signed validated release, its approved campaign and its evidence pack.P1Record to sign: REC-157-1: Release packet per scope: configuration, evidence, limitations, owner, date, rollback (quality manager)
VERA-REQ-158OperationsOperations shall expose per-lane storage growth, transfer lag, queue depth, checksum reconciliation, package health, reference age, QC state and release backlog.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeThe operations counts and readiness show backlogs, queue ages, reconciliation and QC state, as counts only.P1
VERA-REQ-159OperationsRunbooks shall cover instrument workstation failure, edge loss, corrupted transfer, wrong sample, expired reagent, reference withdrawal, model rollback and KG release reversal.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeThe runbooks cover analyser failure, listener loss, a corrupted transfer, a wrong sample, a withdrawn reference and a release reversal; the release reversal is tested.P1Record to sign: REC-159-1: Released runbooks and downtime procedure (go-live B16) (laboratory director)
VERA-REQ-160GovernanceEvery unresolved interface, license, safety, regulatory, data-access or validation question shall remain a visible release gate with owner and due date.Operating Bible v3.0, A.1.4 Safety, simulation, learning, BioSafe and operationsIn launch scopeEvery open item stays visible: unsigned tables hold results with the reason, and the fix list names an owner for each item.P1Record to sign: REC-160-1: Open-items register with owner and due date, reviewed weekly (quality manager)
VERA-REQ-161QCVERA shall evaluate each clinical analyser's QC under the laboratory director's signed rules (1-2s warns; 1-3s, 2-2s, R-4s, 4-1s and 10x reject) and shall not release a result whose QC run or calibration is rejected, missing, expired or not evaluable.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (go-live #19, B10; F26).P0
VERA-REQ-162Critical valuesA result at or beyond a signed critical limit shall be held until a laboratory member logs a call to the clinician with read-back, on the signed escalation clock; a failed attempt shall not clear the hold.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (go-live #20; F25).P0Record to sign: REC-162-1: Critical-result policy and escalation chain, signed (config/lab/critical-escalation.v1.json) (laboratory director)
VERA-REQ-163HoldsA result that fails a signed delta check against the same person's earlier released results, or carries an analyser flag outside the signed releasable list, shall be held for a reviewer's recorded decision.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (go-live #20; F27).P0
VERA-REQ-164ReviewEach clinical result shall be approved or rejected, as it stands in the bundle, by a pathologist or the laboratory director who did not submit the run; a result changed since its review shall need a new review.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (go-live #24, C2).P0
VERA-REQ-165ReleaseWhoever reviewed a result of a bundle, or recorded or alone confirmed the QC that clears it, shall not release it, and the database shall refuse such a release.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (F52; F26).P0
VERA-REQ-166ReleaseOne failed, held or rejected result shall not hold the run's other results; each result left out shall be recorded with the reason, and the report shall say what was not reported.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (F30).P0
VERA-REQ-167RangesEach clinical result shall carry the standard range the signed table gives for the person's recorded sex and age at collection, with the signed rule for other, unknown or unrecorded sex, and the report shall print the range VERA assigned.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (go-live #21, B11; F30).P0Record to sign: REC-167-1: Standard ranges verified in 20 reference individuals of each sex (CLSI EP28), signed in config/lab/reference-intervals.v1.json (laboratory director)
VERA-REQ-168Personal rangesVERA shall read each result against the person's own range once the signed minimum of earlier results exists, keep the standard range beside it as a separately labelled value, never draw a personal range wider than the standard range, and say why it stays on the standard range otherwise.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (personal ranges; VERA-REQ-021, 140, 141).P0
VERA-REQ-169OrdersA clinical result shall be released only for a vial of an order that names an ordering clinician with a DoH licence, carries the collection and data-sharing consent references (genetic testing where ordered), is for an adult, and passed the pre-analytical gate under the signed limits.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (go-live #22, B3, B4).P0
VERA-REQ-170IdentityA primary vial shall be recorded only after the person is identified against the identity vault, and the vial shall keep the method, result, checker and time of the check, never the document number, name or birth date.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (F24).P0
VERA-REQ-171IdentityThe analyser's sample reference shall be the vial's barcode, the run's subject the vial's order's subject and the specimen coding the vial's type; a mismatch shall hold every result of the run.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (review fix (wrong-patient safety net)).P0
VERA-REQ-172ReportThe lab report (signed PDF, HL7 ORU^R01 to the clinic, FHIR) shall carry every item HAAD CLS p13 lists, be authorised only by a registered pathologist or the laboratory director independent of every run, and be refused while a report table is unsigned.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (go-live #26, C3; F27).P0Record to sign: REC-172-1: Report content check against HAAD CLS p13 on the first scope's reports (laboratory director)
VERA-REQ-173CorrectionsA released result shall be corrected only by revoking its release with a printable reason and issuing a corrected report revision that withdraws it and keeps the original, with a notice to the clinic read back for a critical result; the clinic and Malaffi messages shall retract it under the identifiers it was first sent with.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (F28; F31).P0
VERA-REQ-174MalaffiEach clinical release shall queue its one Malaffi result message, built and checked (LOINC per result, DSSID section, range, UCUM unit), and sent at most once by the one sender, also when two dispatchers run at once or a release is replayed.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (go-live #27, C4).P0Record to sign: REC-174-1: Malaffi test message accepted (go-live C4) (Prepaire engineering, with the partners KERNEL and Malaffi (go-live plan C4))
VERA-REQ-175Patient graphOnly a released result shall be written to the person's knowledge graph, sealed under that person's own data key, for the subject of the vial's order; a revoked release shall be superseded in the graph.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (ADR 0061; docs/PATIENT_KG.md).P0
VERA-REQ-176AnalysersThe analyser listener shall capture each copied stream byte for byte into a sealed, hash-chained message log and send nothing to an analyser, which a packet capture shall show.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (go-live #14, B7, B13).P0
VERA-REQ-177AnalysersVERA shall keep KERNEL's results as a separate stored copy and compare it every day with its own reading of the analysers, listing every missing, extra or altered result; the shadow gate shall pass only on 7 consecutive clean days under signed limits.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (go-live #13, #17, B13, B14).P0Record to sign: REC-177-1: Seven-day gap-free capture and 20-day parallel-run reports, signed (PQ) (laboratory director)
VERA-REQ-178AnalysersVERA shall host an analyser's link only after the laboratory director's recorded takeover decision for exactly that analyser, send only orders built from VERA's own conformant orders, and stop at once when the decision is switched off.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (docs/golive/analysers.md sections 6 and 7).P0Record to sign: REC-178-1: Takeover decision per analyser, recorded in VERA by the laboratory director (laboratory director)
VERA-REQ-179AccessOn a copy with a database, every record page and record API shall need a signed-in member of staff with one of its roles (sessions of at most 15 minutes; a service identity never opens a page); the open demonstration copy, which has no database, shall show only invented people and need no sign-in.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (F01).P0
VERA-REQ-180Partner portThe partner port shall stay off on a copy with a database until it is switched on; a partner shall be approved by two different administrators before a credential is issued; partner data shall be read through the signed catalogue; sandbox and real subjects shall never mix.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (docs/partners).P0
VERA-REQ-181Oracle HealthVERA shall open inside a hospital's Oracle Health record only for a registered, switched-on tenant (SMART on FHIR with the brief's scopes), use a pseudonymous subject per record and patient, file its summary under the hospital's own document type, and place no order from the record without the clinician's DoH licence.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (Oracle bolt-on brief, 12 Nov 2026).P0
VERA-REQ-182WearablesA person's wearable readings shall be kept as that person's own records, each with its device and day, read against the person's usual range on that device only after 14 days, and shall never change a laboratory result or its standard range.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (wearables brief, 13 Nov 2026).P0
VERA-REQ-183CKMThe cardiovascular-kidney-metabolic readings (PREVENT, Coachable Health Age, organ age) shall be computed by the published formulas from the person's own results and recorded answers, reproduce the published worked examples, say on every screen that they are research and not a diagnosis, and be held for a real person while the draw intake is unsigned.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (wearables, organ age and CKM brief, 13 Nov 2026).P0
VERA-REQ-184ValidationEvery review, critical call, release and lab report shall record the VERA build that applied the gates; on an operational copy only a numbered release listed in the laboratory director's signed table of validated releases shall approve, release or report a clinical result.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (F53).P0Record to sign: REC-184-1: Table of validated VERA releases, signed (config/lab/validated-vera-releases.v1.json) (laboratory director)
VERA-REQ-185ValidationOn an operational copy a clinical run shall be released only when an approved validation campaign for its analyser and assay protocol version recorded the running VERA image digest, release and clinical table digests, and they match; any difference shall refuse the release.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (F02 step 4; F53).P0Record to sign: REC-185-1: Approved validation campaign per scope, with its decision (laboratory director)
VERA-REQ-186TablesWhile a clinical rule table (catalogue, intervals, QC rules, critical values, delta checks, analyser flags, order rules, report and Malaffi tables) is unsigned, or signed with the demonstration sign-off for a real subject, every clinical result it governs shall be held.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (go-live #16, #19 to #22).P0
VERA-REQ-187AuditEvery decision and record of the result path shall be written to an append-only, hash-chained trail that the database protects against update, delete and truncate, and that a verifier re-walks.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (F21; go-live #9).P0
VERA-REQ-188RetentionEvidence and health records shall be kept at least 25 years (9,132 days) under the object store's COMPLIANCE lock, and an operational copy shall refuse less.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (F12; Federal Law 2/2019 Art. 20).P0
VERA-REQ-189PrivacyNo patient data shall leave the node, the site or the signed UAE allow-list, and operations tools shall answer counts, states and codes only.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (F45; go-live #11).P0
VERA-REQ-190ConsentA withdrawn consent shall stop the uses it covered from the moment it is recorded: twin access for data sharing, and the person's unsent Partners events for affiliate reporting.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (F44).P0
VERA-REQ-191KeysEvery key shall be configured once per copy, be the same in every process (checked by fingerprint), never be kept on a container's temporary disk, and never be printed.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (F08).P0
VERA-REQ-192OperationsHealth shall name the build, mode, key sources and governed configuration, readiness shall check the database and the evidence store, and a stopped process or a backlog shall raise an alert that reaches a named person.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (F11).P0
VERA-REQ-193ChangeEvery database migration shall be additive, so that the version before can run on the newer database and a release can be rolled back.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (PLAT-02; docs/golive/schema.md).P0
VERA-REQ-194ValidationEvery in-scope requirement shall be traced to at least one passing test or a signed record, and the trace matrix and the JUnit report shall be kept with the commit and image digests for 7 years.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (F02 step 3).P0Record to sign: REC-194-1: CI job that keeps the trace matrix and JUnit report for 7 years (docs/validation/README.md) (IT (release engineering))
VERA-REQ-195ValidationEach installation shall have an IQ record (image digests, configuration and table hashes, migrations, keys present) and an OQ record (acceptance and refusal tests on synthetic samples through the release path), written by VERA's scripts for the laboratory director to sign.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (F02 step 4).P0Record to sign: REC-195-1: Signed IQ and OQ records per installation (laboratory director)
VERA-REQ-196TablesA clinical table shall count as signed for the release of a clinical result only when its exact content has a current attested approval by the laboratory director, and by the quality manager as well where the plan asks for both, from two different people; an approval of other content shall never count, and a revocation shall take effect at once.Added since the Operating Bible v3.0In launch scopeA launch function built since the Operating Bible (F18).P0Record to sign: REC-196-1: Attested approvals of every clinical table's content by the laboratory director and, where the plan asks for both, the quality manager (POST /api/v1/governed-tables/attestations) (laboratory director)