VERA™ by Prepaire: validation master plan
| Document | VERA-VAL-MP-001, version 0.3 (draft) |
| Date | 16 November 2026 |
| Owner | The quality manager of IRENA's clinical laboratory, named on the signed approval (section 10) |
| Approver | The laboratory director, named on the signed approval (section 10) |
| Status | Draft for review. Not approved. Nothing in it is signed. |
| Needed | Approved before the first volunteer sample (week of 23 November 2026, go-live plan 3.1 item A2) |
1. Purpose and basis
This plan says how VERA by Prepaire is validated before it reports a patient result at IRENA's clinical laboratory, and after every change. It covers VERA's part of each scope: capturing the analyser's result, translating it, checking it, releasing it after review, reporting it and sending it on. The analysers' own analytical verification (precision, trueness, method comparison, reportable range, carryover, interference, measurement uncertainty and each standard range, checked in 20 reference individuals of each sex) is the laboratory's work on the analysers and is planned separately (go-live plan checklist B1, evidence pack item 4). VERA's validation complements it and never replaces it.
The basis:
- ISO 15189:2022, clause 7.6.3 a). The laboratory information system is verified for its functions by the laboratory before it is introduced, and every change to it, its configuration included, is authorised, documented and validated before it is put into use. VERA is that system for every scope it releases.
- The go-live plan's own rule (VERA in production, 9 November 2026, sections 1 and 7): no test is reported to a patient, by KERNEL or by VERA, before the laboratory director has signed its verification, and VERA reports nothing until its evidence pack for that scope is signed.
- DoH's request to validate each analyser and test before patient testing (go-live plan 6.2, REG A3).
- ISO 22367:2020 for the risk assessment of the result path (docs/validation/risk-assessment.md).
- The requirements of the Prepaire VERA Platform Operating Bible v3.0, Appendix A.1, and the launch functions added since (section 5).
2. Scope
2.1 What one scope is
A scope is the unit that is validated, signed and released (go-live plan 7.4). It is exactly:
- one analyser: its register entry (asset id and serial) and its middleware and firmware versions;
- one assay protocol version on that analyser;
- one VERA platform image and one VERA worker image, each by digest (
sha256:...), and the numbered release they were built from (vera-YYYY.MM.DD-nand its commit); - one pipeline definition (its SHA-256);
- the clinical tables VERA applies, each by its content digest (
src/lib/validation/clinical-tables.ts): the catalogue (code map, units, specimens and tubes), the standard ranges by sex and age, the QC rules and targets, the critical limits and their call clock, the delta checks, the releasable analyser flags, the order rules, the report's laboratory facts, signatories and coded references, the Malaffi section map and profile, KERNEL's code crosswalk, the unit conversions, the Sysmex host codes, the parameters of the ingestion gate, the biological variation and the genotype-specific starting ranges behind personal ranges, and the calculated indices. The other tables of F18's list (the PQ limits, the ingestion framework's descriptive tables, the Olink and report-import terms, the analyser register's side-by-side block) are governed and attested too, but are not part of a release's scope.
VERA enforces the scope itself. A validation campaign records it (POST /api/v1/validation/campaigns with scope, which
must name the worker image and the pipeline definition as well as the platform image and the tables); a campaign whose
scope lacks any of them, or that was opened on a copy that is not a numbered release, validates no scope. VERA scores
the campaign's trials (its evidence decision); the laboratory director approves that decision in VERA
(POST /api/v1/validation/campaigns/{campaignId}/approval), and may withdraw the approval. On an operational copy the
release route refuses a clinical run unless the copy and the run are in a scope whose latest word is the director's
current approval: a withdrawal, or a later campaign of the same scope that VERA rejected, ends it at once, also for a
release already being prepared (409 VALIDATION_SCOPE_NOT_APPROVED or VALIDATION_SCOPE_MISMATCH, F02 step 4). A build
not listed in the laboratory director's table of validated releases cannot release either (409 VERA_BUILD_NOT_VALIDATED,
F53). A clinical table counts as signed for the release, and for report issue, only when its exact content has its
attested approvals in VERA's database: the laboratory director's, and the quality manager's as well for the catalogue,
the standard ranges, the QC rules, the critical values and the delta checks, by two different people; the table of
validated releases needs the laboratory director's attestation too (POST /api/v1/governed-tables/attestations;
409 TABLES_NOT_ATTESTED otherwise, F18). The file's own review fields are display only.
2.2 The scopes, in order
The order follows go-live plan 7.2. The serials, software and firmware versions are read from the nameplates (checklist A7) and are not yet known.
| Order | Analyser | Tests | Asset id and serial |
|---|---|---|---|
| 1 | Roche cobas pure (c 303 and e 402) | Chemistry and immunoassay of GenetiQ Check | [to confirm] |
| 2 | Roche cobas i 601, only if founder decision 7 keeps the hormones there | Hormones | [to confirm] |
| 3 | Sysmex XR-1000 | Blood count (20 markers) | SRA-SXA-000001 in the register; serial [to confirm] |
| 4 | Sysmex urine line (UC-3500, UF-5000, UD-10 through U-WAM) | Urinalysis (20 markers) | [to confirm] |
The stool track (MGI DNBSEQ-T20) is validated on its own plan and is not part of this one (go-live plan 3.3, S1 to S8).
2.3 What VERA™ does in a scope, and what is validated
| Step | What VERA does | Validated by |
|---|---|---|
| Order and vial | Accepts an order only with a licensed ordering clinician, consents and an adult; identifies the person at the draw; checks the vial's pre-analytical limits | OQ-21; VERA-REQ-169, 170 |
| Capture | Keeps KERNEL's results as a separate copy; reads a copy of the analyser's own messages and never sends to it; after the laboratory director's recorded takeover, hosts that analyser's link | OQ-16, OQ-17, OQ-24; PQ shadow stage |
| Translate | Maps each result to the catalogue's marker, LOINC code and UCUM unit, and assigns the standard range for the person's sex and age | OQ-14, OQ-22; PQ parallel run |
| Check | Westgard QC and calibration lockout, critical values held until the call is logged, delta checks, analyser flags, unsigned tables hold every result | OQ-04, OQ-05, OQ-13, OQ-18 |
| Review and release | Each result approved by a pathologist or the laboratory director who did not submit the run; released by a second person, per result; only by a validated release in a scope the laboratory director approved; the clinical routes keep the sign-in gate | OQ-01 to OQ-03, OQ-08 to OQ-11, OQ-27 to OQ-29 |
| Report | Signed PDF, HL7 ORU^R01 to the clinic and FHIR, with every item HAAD CLS p13 lists; corrected reports keep the original | OQ-19, OQ-20 |
| Messages | One Malaffi message per release, sent at most once by the one sender | OQ-12, OQ-15 |
| Record | Writes only released results to the person's encrypted knowledge graph; reads each result against the person's own range once enough results exist | VERA-REQ-146, 168, 175 |
2.4 Interfaces in scope
| Interface | Direction | In this plan |
|---|---|---|
| Analyser listener (ASTM E1381/E1394, HL7 MLLP) | Copy of the analyser link, read only | Yes, every scope |
| Analyser host mode | VERA answers the analyser and sends its orders, per analyser, only after the laboratory director's recorded takeover | Yes, per analyser, before its takeover (docs/golive/analysers.md) |
| KERNEL | VERA reads KERNEL's finished results (the shadow copy); KERNEL is recommended to keep the accession and the patient's identity at launch (founder decision 5, a recommendation due 23 November 2026 [to confirm]) | Yes |
| Malaffi | VERA builds and checks the result message; one sender sends it (recommended: KERNEL, triggered by a VERA release; founder decision 6, a recommendation due 23 November 2026 [to confirm]) | Yes, before the first VERA release (checklist C4) |
| Clinic report | Signed PDF, HL7 ORU^R01, FHIR Bundle, handed over with the hand-over recorded; no VERA transport sends them at launch [to confirm, go-live plan C3] | Yes |
| Hospital orders and Oracle Health (SMART on FHIR) | Orders in; VERA's summary filed in the hospital record | Validated before a hospital tenant is switched on [to confirm with the hospital] |
| Partner port | Partner data in, results out, two administrators approve each partner | Validated before the port is switched on for a partner |
Interfaces outside VERA (the analysers' own interfaces, KERNEL's Malaffi sender, the hospital's record) are qualified by their owners; VERA's records name the versions it was validated against.
3. Roles
| Role | What the role does here | Cannot |
|---|---|---|
| Laboratory director, named on each record they sign, and by their own sign-in on each approval and attestation they make in VERA | Approves this plan, the risk assessment, the requirements register's scope decisions, each IQ, OQ and PQ record; attests each clinical table and the table of validated VERA releases in VERA; approves each campaign's scope in VERA and withdraws it when it no longer holds; signs the evidence pack | Approve a record they performed; approve a campaign whose trials they posted |
| Quality manager, a different person from the laboratory director, named on each record they sign, and by their own sign-in on each attestation they make in VERA | Owns this plan and every validation document and record; reviews each record; attests the tables the plan gives both people; keeps the change records and the deviation log; adds signed records to config/validation/records.v1.json | Be the laboratory director; review a record they performed (the laboratory director, or a second reviewer, reviews it [to confirm]) |
| Tester: IT, a laboratory scientist or the quality manager, never the developer | Runs the IQ and OQ scripts and the PQ steps, and is named as performedBy in each record (the tester's own statement: the scripts run where there is no sign-in, and the signed page binds it) | Run a test of their own code |
| Developer (Prepaire engineering, with Claude) | Drafts these documents; builds and tests the code; writes the trace matrix | Sign, approve or run a qualification record |
| IT and release engineering | Deploy the release, read the image digests, set VERA_IMAGE_DIGEST and VERA_WORKER_IMAGE_DIGEST, give the platform a database role that owns no table, wire the CI job that keeps the records (docs/validation/README.md) | Change the code or the tables |
| Executive go-live authority (the founder) | Signature 8 of the evidence pack | Sign a laboratory record |
The evidence pack carries eight signatures (laboratory director, quality manager, clinical safety officer, information security officer, privacy and legal, LIM product owner, KG owner, executive go-live authority) and two VERA attestations made by different named people (go-live plan 7.4 item 14). Several of these people are not yet named (go-live plan decision 3).
4. Stages and their exit
| Stage | What happens | Exit (signed by the laboratory director) | When (go-live plan 7.2) |
|---|---|---|---|
| Plan | This plan, the risk assessment and the acceptance limits | Signed before the first volunteer sample | weeks 1-2, 16-29 Nov |
| Requirements | The register (config/validation/requirements.v1.json) and its trace; CI keeps the trace matrix of every commit | Register signed; the trace matrix of the release passes | with each release |
| IQ | scripts/validation/iq.ts on each node that will release, for each release | Signed IQ record | before OQ |
| OQ | scripts/validation/oq.ts: 29 cases, 27 acceptance and refusal cases on synthetic samples through the release path and 2 asked of the installed node | Signed OQ record citing the IQ record | before PQ counts |
| PQ | Shadow (7 days gap-free) and parallel run against KERNEL (20 working days), docs/validation/pq-protocol.md | Signed PQ record; the campaign's evidence decision at conditional or better, approved by the laboratory director in VERA | weeks 5-11, Dec to Jan |
| Evidence pack | One per scope (go-live plan 7.4) | 8 signatures and 2 attestations; the release listed in config/lab/validated-vera-releases.v1.json and that table attested; the director's approval of the campaign | week 13, Feb |
IQ, OQ and PQ are repeated on every change that the change rule (section 7) says needs them.
5. Requirements and traceability
- The register:
config/validation/requirements.v1.json. VERA-REQ-001 to 160 are Appendix A.1 of the Operating Bible v3.0, word for word; VERA-REQ-161 onward are the launch functions built since (QC lockout, critical hold and call clock, per-result review, release by a second person, release per result, the standard range by sex and age, personal ranges, orders, identity, the HAAD CLS report, corrected reports, Malaffi send-once, the patient graph, the listener, the KERNEL shadow, host mode, sign-in on record pages, the partner port, Oracle Health, wearables, CKM, the validated build and scope, the attested table signatures, and the operations requirements). Each requirement is in or out of launch scope with its reason; 108 of the 196 are in, 3 of them in part (said in words). The scope decisions are proposals for the laboratory director. - The trace:
config/validation/trace.v1.jsonlinks each in-scope requirement to the tests that prove it; a test whose title names a requirement id is linked too.src/lib/validation/trace-matrix.test.tsfails the build while an in-scope requirement has no test. Where the tests show only part of what is in scope, the trace map says what they do not show (partial), and the matrix reports the requirement as shown in part, never as passed. Where only part of a requirement's text is in launch scope, the register says which part in words (launchPart), and "passed" is for that part. - Signed records: where a test cannot show a requirement (an SOP, a network record, a signed table), the register
names the record and its owner, and
config/validation/records.v1.jsonlinks it once signed. - The trace matrix:
scripts/validation/trace-matrix.tsjoins a test run with the register and writes the matrix and a JUnit report, with the commit and the image digests; CI keeps them for 7 years (docs/validation/README.md). Its--strictmode, which also waits for every signed record and every requirement shown only in part, is the gate for a release that will report patients.
6. Acceptance limits
Proposed in go-live plan 7.3. The laboratory director decides each.
| Check | Pass rule | How VERA shows it |
|---|---|---|
| Capture completeness | 0 missing or extra messages over 7 consecutive days, against analyser and KERNEL counts | The shadow gate and the daily reconciliation report (PQ) |
| Byte fidelity | Replayed fixtures give identical SHA-256 | OQ-17; the listener's replay check |
| Value, unit, flag | 100% identical to the instrument record, or converted exactly as the signed table says | The side-by-side and parallel-run comparisons (PQ) |
| Standard range assignment | 100% of persons get the standard range for their sex and age; the report matches the system | OQ-14 |
| QC lockout | 100% of seeded failures blocked | OQ-04, OQ-18 |
| Transfer of a standard range | No more than 2 of 20 per group fall outside (CLSI EP28) | The laboratory's verification, recorded through VERA's EP28 checker |
| Evidence decision | Conditional at 71 effective clean trials (lower bound 0.95); approved at 367 (0.99); each specimen counts once per kind, operator, lot and site | The campaign's evidence decision under PREPAIRE-VENDOR-EVIDENCE-POLICY@1.0.0, then the laboratory director's approval in VERA |
"Conditional" is how much evidence VERA's own copying and translating has earned; it is not a lower bar for the patient. A result released under it has passed every limit above (go-live plan 7.1).
7. The change rule
Every change that can reach a node is authorised, documented and validated before it reaches the node. This covers the
code, the images, the clinical tables and every other file under config/, the database migrations, the dependencies,
and the deployment settings of an operational node.
- Record. The change has a change record: what changes, why, and its effect on the risk assessment (which hazards, which controls). The quality manager owns the record.
- Authorise. The quality manager authorises it; a change to a clinical table is also signed by the laboratory director in the table itself and attested in VERA for its new content (and both people where the plan asks for both: catalogue, ranges, QC rules, critical values, delta checks). An edit after an attestation is unsigned until it is attested again.
- Build and test. The change is made on a branch, every migration stays additive
(
scripts/release/check-migrations.ts), and the trace matrix of its commit passes in CI. - Validate. The release is cut (
vera-YYYY.MM.DD-n) and its scope sheet written (iq.ts sheet). On each node: IQ, then OQ. The quality manager and the laboratory director decide, with the risk assessment, whether the PQ is repeated in full, in part or not (proposal: in full when a clinical table, the catalogue, the capture or the release path changes; the OQ alone for a change that touches none of them; [to confirm] by the laboratory director). - Approve. The laboratory director adds the release to
config/lab/validated-vera-releases.v1.jsonand attests that table's new content in VERA, and approves in VERA the campaign that records the new scope. VERA enforces this step: a release outside the attested table or an approved scope cannot release a clinical result, and a changed clinical table changes the scope's digest. The director withdraws an approval in VERA when its scope no longer holds. - Keep. The change record, the trace matrix, the IQ, OQ and PQ records and the approvals are kept together with the commit and the image digests.
There is no emergency path around these steps in VERA. In an emergency the operations switches close the release path, KERNEL keeps reporting, and the change follows the rule before VERA releases again (go-live plan 8.3; downtime and return-to-KERNEL procedures, checklist B16).
8. Deviations
A failed check, a failed test or a discordance is a deviation. It is logged by the quality manager with its root cause (VERA, KERNEL, the mapping or the analyser), its correction and what is re-tested. A record with a failed check is signed only with the deviation and its disposition attached. In the parallel run every discordance is traced to its cause before the run continues and counts against the evidence score (go-live plan 7.1 and 8.2). An identity mismatch, an altered source value, a release without QC or a critical value not called stops the pilot.
9. Documents and records
| Document or record | Where | Kept |
|---|---|---|
| This plan, the risk assessment, the IQ, OQ and PQ protocols | docs/validation/ (under version control) | With the release |
| Requirements register, trace map, signed records, OQ protocol | config/validation/ | With the release |
| Trace matrix and JUnit report of each commit | CI artefacts, then the evidence store | 7 years [to confirm against the laboratory's retention schedule] |
| IQ and OQ records (JSON, the page signed, the test reports) | validation-records/ on the node, then the evidence store | 7 years [to confirm] |
| PQ records: daily shadow reports, side-by-side reports and review sheets, campaign trials and decisions | VERA's database (append-only), the evidence store | With the laboratory's records (25 years for records tied to patient results, F12) |
| Change records and deviation log | The quality manager's document system [to confirm which] | 7 years [to confirm] |
10. Approval
Signing this plan approves it for the validation of VERA's scopes at IRENA. Each signature names the version above.
- Prepared by (developer): Prepaire engineering, 16 November 2026
- Reviewed by (quality manager): name ______________________ date __________ signature ______________________
- Approved by (laboratory director): name ______________________ date __________ signature ______________________
Change history
| Version | Date | Change | By |
|---|---|---|---|
| 0.1 | 15 Nov 2026 | First draft (F02 of the readiness fix list) | Prepaire engineering |
| 0.2 | 16 Nov 2026 | The laboratory director approves each scope in VERA and may withdraw it; a scope names the worker image and the pipeline; the tables of validated releases attested; partial coverage in the trace; founder decisions 5 and 6 as recommendations; roles by role, and no one reviews a record they performed | Prepaire engineering |
| 0.3 | 16 Nov 2026 | The owner and the approver are named where they sign, not in advance; the standard range in VERA's own words (section 1 and the acceptance limits) | Prepaire engineering |